Installation
Turbimax CUS52D
14
Hauser
Pipes made of stainless steel with diameter >DN 300 exhibit hardly any wall effects.
Black plastic pipes with diameter > DN 60 exhibit hardly any wall effects (<0.05 FNU). For
this reason, the use of black plastic pipes is recommended.
6
7
8
9
10
A0030704
9
Orientations for pipes and assemblies
• Install the sensor in such a way that the light beam is not reflected → 9, 14 (item
6).
• Avoid sudden changes in cross-section (item 9). Changes in cross-section should be
gradual and located as far away as possible from the sensor (item 10).
• Do not install the sensor directly downstream from a bend (item 7). Instead position it as
far away as possible from the bend (item 8).
Installation marking
1
A0030820
10
Installation marking for sensor alignment
1
Installation marking
The installation marking on the sensor is aligned opposite the optical system.
‣
Align the sensor against the flow direction.