Operation
Turbimax CUS50D
36
Hauser
3. Compare the measured value with the reference value on the calibration kit.
The function check is positive if the deviation is within the permitted tolerances
Reference tool
Solid state reference, CUS50D kit
Tolerance
± 5%
± 10%
If the measured values of a calibration data record are displayed instead of the raw
values, the measured values may differ due to the calibration, offset or factor.