Principle of an offset

x

Measured value

y

Target sample value

a

Factory calibration

b

Offset calibration

8.1.3  

Cyclic cleaning

For cyclic cleaning, compressed air is the most suitable option. The cleaning unit is either

supplied or can be retrofitted, and is attached to the sensor head. The following settings are

recommended for the cleaning unit:

Type of fouling

Cleaning interval

Cleaning duration

Severe fouling with rapid buildup of deposits

5 minutes

10 seconds

Low risk of fouling

10 minutes

10 seconds

8.1.4  

Signal filter

The sensor is fitted with an internal signal filter function in order to adapt the measurement

flexibly to different measuring requirements. Turbidity measurements based on the principle

of scattered light may have a low signal-to-noise ratio. In addition, disturbances can occur as a

result of air bubbles or fouling, for example.
It is not practicable to even out these disturbances using a high level of damping. This would

counteract the sensitivity of the measured value required in the applications.

Measured value filter

Summary of Contents for Turbimax CUS50D

Page 1: ...Products Solutions Services Operating Instructions Turbimax CUS50D Absorption sensor for turbidity and solids measurements BA01846C 07 EN 03 20 71473151 2020 02 10...

Page 2: ......

Page 3: ...sor 16 5 3 Mounting the compressed air cleaning unit 20 5 4 Post installation check 20 6 Electrical connection 21 6 1 Connecting the sensor 21 6 2 Ensuring the degree of protection 23 6 3 Post connect...

Page 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Page 5: ...Turbimax CUS50D About this document Endress Hauser 5 1 3 Symbols on the device Symbol Meaning Reference to device documentation...

Page 6: ...rticularly suited for use in the following applications Turbidity measurements based on the principle of light attenuation turbidimetry in accordance with EN ISO 7027 Absorption measurements in liquid...

Page 7: ...oducts as defective During operation If faults cannot be rectified products must be taken out of service and protected against unintentional operation 2 5 Product safety 2 5 1 State of the art technol...

Page 8: ...iption 3 1 Product design The sensor features a sensor head with 2 path lengths of 5 mm 0 2 in and 10 mm 0 39 in 1 2 3 4 A0036825 1 CUS50D sensor head 1 Light sources 10 mm 0 39 in 2 Light sources 5 m...

Page 9: ...h compressed air cleaning 3 1 1 Measuring principle The sensor operates on the principle of light attenuation in accordance with ISO 7027 and meets the requirements of this standard It is suitable for...

Page 10: ...ed against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact your supplier or you...

Page 11: ...f the product by affixing to it the mark 4 4 2 Electromagnetic compatibility Interference emission and interference immunity as per EN 61326 1 2013 EN 61326 2 3 2013 NAMUR NE21 2012 4 4 3 ISO 7027 The...

Page 12: ...ax CUS50D 12 Endress Hauser 5 Installation 5 1 Installation conditions 5 1 1 Dimensions 82 7 3 26 12 6 20 5 0 8 292 2 11 5 40 1 57 209 5 8 25 24 5 1 NPT3 4 GI 36 1 4 0 5 13 4 0 53 A0036366 3 Dimension...

Page 13: ...bimax CUS50D Installation Endress Hauser 13 82 7 3 26 12 6 20 5 0 8 292 2 11 5 40 1 57 209 5 8 25 24 5 1 NPT3 4 GI 36 1 4 0 5 13 4 0 53 85 9 3 38 64 2 52 A0036582 4 Dimensions with clamp Dimensions mm...

Page 14: ...tion Turbimax CUS50D 14 Endress Hauser 61 2 4 64 5 2 5 36 9 1 5 4 7 0 19 3 8 1 5 40 1 6 A0036826 5 Dimensions for compressed air cleaning Dimensions mm in Compressed air cleaning 2 bar 29 psi maximum...

Page 15: ...ameter must be at least 50 mm 2 in Install the sensor in places with consistent flow conditions The best installation location is in the ascending pipe item 1 A0036370 7 Direction of flow Align the se...

Page 16: ...nstallation marking align the sensor against the flow direction 5 2 Mounting the sensor 5 2 1 Measuring system A complete measuring system comprises Turbimax CUS50D turbidity sensor Liquiline CM44x mu...

Page 17: ...annel transmitter 2 Turbimax CUS50D turbidity sensor 3 CUA252 flow assembly 4 Direction of flow 5 2 2 Mounting options Installing with CUA120 flow assembly 2 A0036835 10 Installing with CUA120 flow as...

Page 18: ...h CUA252 flow assembly The installation angle is 90 The arrow indicates the flow direction it runs from the 10 mm 0 39 in path to the 5 mm 0 2 in path A0036836 12 Installing with CUA262 flow assembly...

Page 19: ...runs from the 10 mm 0 39 in path to the 5 mm 0 2 in path Installing with CUA451 retractable assembly 2 A0036838 14 Installing with CUA451 retractable assembly The installation angle is 90 The arrow i...

Page 20: ...a way that air bubbles cannot accumulate on it 5 3 Mounting the compressed air cleaning unit Fit the compressed air cleaning unit onto the sensor head to the end stop The nozzle of the compressed air...

Page 21: ...erating Instructions and must follow the instructions contained therein Prior to commencing connection work ensure that no voltage is present on any cable 6 1 Connecting the sensor Connect the sensor...

Page 22: ...o sensor input left or via M12 connector right Connecting the cable shield Cable sample does not necessarily correspond to the original cable supplied 1 2 3 18 Terminated cable 1 Outer shield exposed...

Page 23: ...ble glands Modules are not fully secured The display is not fully secured risk of moisture entering due to inadequate sealing Loose or insufficiently tightened cables cable ends Conductive cable stran...

Page 24: ...issioning 7 1 Function check Prior to initial commissioning ensure that The sensor is correctly installed The electrical connection is correct Before commissioning check the chemical compatibility of...

Page 25: ...erform customer calibrations with up to 10 points Application Formazine Factory calibration for the formazine application is carried out with the formazine turbidity standard Sensor measured values in...

Page 26: ...ct the application that suits your field of application Model name Application Unit Absorption Absorption measurement in any liquid media dosing of flocculant AU OD Formazine Absorption turbidity meas...

Page 27: ...l mg l ppm There is also the additional option of selecting User unit as the unit Here any unit name or a character string can be assigned to the Basic unit e g OD for absorption The system can be ca...

Page 28: ...alibration The measured error between the measured value of the device and the laboratory measured value is too large This is corrected by a 1 point calibration 1 2 3 4 5 6 7 8 1 2 3 4 5 6 x y a b A00...

Page 29: ...point calibration x Measured value y Target sample value a Factory calibration b Application calibration 1 Select a data record 2 Set 2 different calibration points in the medium and enter the corresp...

Page 30: ...lation is performed outside the calibrated operational range gray line The calibration curve must be monotonically increasing The following sample values for the calibration of the CUS50D sensor can b...

Page 31: ...ation process The aim is to achieve the highest possible calibration quality in the shortest possible time frame while taking external conditions into account For calibrations in the field in adverse...

Page 32: ...8 1 2 3 4 5 6 x y f 1 1 a b A0039329 24 Principle of factor calibration x Measured value y Target sample value a Factory calibration b Factor calibration Offset With the Offset function the measured v...

Page 33: ...l Cleaning duration Severe fouling with rapid buildup of deposits 5 minutes 10 seconds Low risk of fouling 10 minutes 10 seconds 8 1 4 Signal filter The sensor is fitted with an internal signal filter...

Page 34: ...ression filter in liquids with a high level of turbidity with a high solids content Air bubbles do not cause the measured value to increase in media of this type and thus cannot be eliminated with the...

Page 35: ...rection on the sensor head B The direction is specified on the calibration kit 4 A B A0036702 Slide the calibration kit A into the final position until the limit stop B Function check The sensor raw v...

Page 36: ...it The function check is positive if the deviation is within the permitted tolerances see 34 Reference tool Solid state reference CUS50D kit Tolerance 5 10 If the measured values of a calibration data...

Page 37: ...ting Solution No display no sensor reaction Power supplied to transmitter Sensor connected correctly Buildup on optical windows Connect mains voltage Connect sensor correctly Clean sensor Display valu...

Page 38: ...The medium in which measurement takes place 10 1 Maintenance tasks 10 1 1 Clean sensor Sensor fouling can affect the measurement results and even cause a malfunction The sensor must be cleaned regula...

Page 39: ...ion are required or if the wrong product was ordered or delivered As an ISO certified company and also due to legal regulations Endress Hauser is obliged to follow certain procedures when handling any...

Page 40: ...62 Weld in flow assembly Product Configurator on the product page www endress com cua262 Technical Information TI01152C Flexdip CYA112 Immersion assembly for water and wastewater Modular assembly syst...

Page 41: ...anywhere on the ground on the coping stone on the wall or directly onto railings Stainless steel version Product Configurator on the product page www endress com cyh112 Technical Information TI00430C...

Page 42: ...US50D Connection 6 mm 0 24 in Pressure 1 5 to 2 bar 21 8 to 29 psi Materials POM PE PA 6 6 30 glass fiber Order number 71395617 Compressor For compressed air cleaning 230 V AC order number 71072583 11...

Page 43: ...0 g l Product loss 0 to 100 1000 Measuring range with solids content For solids the achievable ranges depend very much on the media that are actually present and may differ from the recommended operat...

Page 44: ...ed for calibration 13 2 3 Drift Working on the basis of electronic controls the sensor is largely free of drifts Formazine drift 0 04 per day for 2000 FAU Absorption drift 0 015 per day for 5 OD 13 2...

Page 45: ...tion Installation 13 5 2 Weight Cable length Plastic sensor Metal sensor Metal sensor with clamp 3 m 9 84 ft 0 46 kg 1 5 lbs 1 15 kg 2 54 lbs 1 21 kg 2 67 lbs 7 m 23 ft 0 68 kg 1 5 lbs 1 37 kg 3 81 lb...

Page 46: ...put 43 Installation 12 16 Installation conditions 12 M Maintenance 38 Marine 11 Measuring principle 9 Measuring system 16 Mechanical construction 45 Multipoint calibration 27 N Nameplate 10 O Offset 3...

Page 47: ......

Page 48: ...www addresses endress com 71473151 71473151...

Reviews: