Trouble-shooting
39
9.4
Spare parts
Fig. 19 shows the components of the analyser. Please, take the spare parts order numbers from the
following sections.
9.4.1
Spare parts for sample and reagent transportation
Flushing does not start
Time of 1st flushing not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired or 0
h
Change parameters.
Leak on
Photometer
Device or discharge
blocked
Remove blockage.
Service
Blockage, deposits in
device
Water hardness
Lime deposits can by flushing with 5% hydrochloric acid .
If necessary, dose EDTA in sample stream to prevent sedimentation (do not use EDTA with CA71HA!).
Not enough sample
conditioning
Shorten cleaning intervals of sample conditioning..
Error
Possible cause
Tests and / or corrective measures
C07-CA71XXx-00-08-xx-004.eps
Fig. 19:
Overview parts
A
B
C
P
Sample input two-channel version
Standards and reagent canisters
Collecting vessel
Sample
R
R1
S
Cleaner canister
Reagent 1 canister
Standard solution canister
Item
Spare part
Order number
120
Hose made of Norprene, 1.6 mm
51504116
121
Hose made of C-Flex, 3.2 mm (permeate inlet and overflow with sample collector)
51504114
122
Hose made of C-Flex, 6.4 mm
51504115
A
R
S
R1
P
B
B
C
120
131
130
123
161
160
120
200
120
122
142
144
140
155
131
134
130
141
143
121
135
135
121
170/171
154
P
P
P
Summary of Contents for Stamolys CA71NO
Page 2: ......
Page 54: ...54...
Page 56: ...www endress com worldwide BA355C 07 en 01 03 Printed in Germany FM SGML 6 0 DT 51517627...