XA02750C
SS2100 TDLAS Gas Analyzer
8
Hauser
Personnel responsibility
Mitigating Risks
Refer to the instructions for each situation listed below to mitigate associated
risks.
Exposure to toxic gas (H
2
S)
Follow the procedure below if there has been any suspected leak from the
sample system and accumulated SCS enclosure.
1.
Purge the SCS enclosure to remove any potentially toxic gas.
2.
Test the H
2
S levels of the SCS enclosure using the port from the safety
purge kit to ensure the purge has cleared any toxic gas.
3.
If no gas leak is detected, open the SCS enclosure door.
Electrocution hazard
1.
Shut off power at the main disconnect external to the analyzer.
2.
Open enclosure door.
If service must be performed with power engaged:
1.
Note any live electrical components and avoid all contact with them.
2.
Only use tools with a safety rating for protection against accidental
contact with voltage up to 1000V (IEC 900, ASTF-F1505-04, VDE
0682/201).
Explosion/fire hazard
Any work in a hazardous area must be carefully controlled to avoid creating
any possible ignition sources (e.g., heat, arcing, sparking, etc.). All tools must
be appropriate for the area and hazards present. Electrical connections must
not be made or broken with power on (to avoid arcing).
The safety of the analyzer is the responsibility of the installer
and the organization he/she represents.
Follow all safety protocols governing toxic gases and
potential leaks.
Complete this action before performing any service that
requires working near the main input power or
disconnecting any wiring or other electrical components.