Solitrend MMP60
Commissioning
Hauser
17
10
20
30
40
50
100
200
300
400
500
600
700
60
t [ps]
H [%]
12
8
14
15
9
13
A0037432
7
Calibration curve set A (Cal.8, Cal.9, Cal.12, Cal.13, Cal.14, Cal.15)
H
Gravimetric moisture; %
t
Radar transit time; picoseconds
8
Cal.8, Lignite
9
Cal.9, Basic calibration
12 Cal.12, Sewage sludge
13 Cal.13, Cereals (linear)
14 Cal.14, Air/water 0 to 100 %
15 Cal.15, Raw data calibration (
1
/
10
of average radar transit time)
The graphics show the linear calibration curves (Cal.1 to Cal.15) for various materials that are
saved and can be selected in the device. The gravimetric moisture (H) is indicated as a
percentage on the y-axis, and the associated radar transit time (t) in picoseconds is indicated
on the x-axis. The radar transit time is displayed simultaneously with the moisture value
during moisture measurement. In air, the devices measure approx. 60 ps radar transit time,
and 1
000 ps in water.
8.4
Settings
8.4.1
Material calibration
In the
Material calibration
menu item, the necessary calibration can be selected via the
optional remote display, depending on the application. In this way, one device can cover a
range of different applications.