Commissioning

Solitrend MMP60

14

Hauser

Moisture range in %

• Maximum: e.g. 20 %

• Minimum: 0 %

Temperature range in °C

• Maximum: 100 °C, this also applies for the high-temperature version.

• Minimum: 0 °C

Conductivity in mS/cm

• Maximum 20 mS/cm

• Minimum 0 mS/cm

The devices can measure the conductivity depending on the device type and the

moisture. The output is set to 0 to 20 mS/cm at the factory.

8.2  

Operating mode

The device configuration is preset at the factory before the device is delivered. This device

setting can then be optimized to suit the process.

Measure mode and parameters:

The following device settings can be changed

• Measure mode C - Cyclic (default setting for devices with cyclic measurement).

• Average time, reaction speed of the measured values

• Calibration (when different materials are used)

• Filter function

• Precision of a single value measurement

Each of these settings is retained even after the device is switched off, i.e. the setting is

saved to the non-volatile memory of the device.

8.2.1  

Operating mode

The device is supplied from the factory with the 

CH

 mode for applications in the construction

industry, and with the 

CA

 mode for general process applications. 6 different operating modes

are available in the 

C

 measuring mode, depending on the application.

CS

 mode (Cyclic-Successive)

For very short measurement cycles in the seconds range (e.g. 1 to 10 s) without averaging

and without filter functions, and with up to 100 measurements per second internally and a

cycle time of 250 ms at the analog output.

CA

 mode (Cyclic Average Filter)

Standard averaging for relatively fast but continuous measurement processes, with simple

filtering and an accuracy of up to 0.1 %. The 

CA

 operating mode is also used to record raw

values, without averaging and filtering, to be able to subsequently analyze the measured

data and determine the optimum operating mode.

CF

 mode (Cyclic Floating Average with Filter)

Floating average for very slow and continuous measurement processes, with simple filtering

and an accuracy of up to 0.1 %. Suitable for applications on a conveyor belt etc.

Summary of Contents for Solitrend MMP60

Page 1: ...tructions they are not a substitute for the Operating Instructions pertaining to the device For detailed information refer to the Operating Instructions and other documentation Available for all devic...

Page 2: ...Solitrend MMP60 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Page 3: ...7 4 2 Product identification 7 4 3 Manufacturer s address 8 4 4 Storage transport 8 5 Mounting 8 5 1 Mounting requirements 8 5 2 Mounting the device 8 5 3 Post mounting check 9 6 Electrical connection...

Page 4: ...ollowing requirements Trained qualified specialists must have a relevant qualification for this specific function and task Personnel must be authorized by the plant owner operator Be familiar with fed...

Page 5: ...t Observe the limit values in the Technical data Incorrect use The manufacturer is not liable for damage caused by improper or non intended use Clarification of borderline cases With regard to special...

Page 6: ...ements has been tested and left the factory in a condition in which it is safe to operate It meets the general safety standards and legal requirements It also complies with the EU directives listed in...

Page 7: ...ditions is not met please contact the manufacturer s sales office 4 2 Product identification The device can be identified in the following ways Nameplate specifications Extended order code with breakd...

Page 8: ...lace of installation to ensure that the material flow and therefore the bulk density over the measurement surface is constant The measuring field of the device must be completely covered by material a...

Page 9: ...ns Unit of measurement mm in 5 2 1 Mounting the housing with the remote electronics module The housing with the remote electronics module can be mounted with two screws M5 163 6 42 52 2 05 8 0 3 1 A00...

Page 10: ...ured 6 Electrical connection 6 1 Connecting requirements 6 1 1 Cable specification Connecting cables with a pre assembled 10 pin socket on the device side are available in different standard lengths 4...

Page 11: ...WH G RS485 B must be enabled Wire color brown BN C IMP Bus RT Wire color gray GY pink PK see figure below J IMP Bus COM Wire color blue BU red RD see figure below K 2nd analog positive Wire color pin...

Page 12: ...g output The device can be set to 0 to 20 mA or 4 to 20 mA The current output can also be set inversely to 20 to 0 mA or 20 to 4 mA for special controllers and applications The analog outputs can be s...

Page 13: ...10 0 to 20 or 0 to 30 8 1 1 Possible settings There are several possible settings for the analog outputs Analog outputs Options 0 to 20 mA 4 to 20 mA The current output can also be set inversely for...

Page 14: ...setting is saved to the non volatile memory of the device 8 2 1 Operating mode The device is supplied from the factory with the CH mode for applications in the construction industry and with the CA mo...

Page 15: ...o discharge hatch The CH mode performs filtering automatically This allows drip water that forms in the silo to be filtered out of the measured value for example 8 3 Calibration curve set A for genera...

Page 16: ...6 Calibration curve set A Cal 1 Cal 2 Cal 3 Cal 4 Cal 5 Cal 6 Cal 7 Cal 10 Cal 11 H Gravimetric moisture t Radar transit time picoseconds 1 Cal 1 Universal sand gravel grit 2 Cal 2 Sand 1 6 3 Cal 3 S...

Page 17: ...urves Cal 1 to Cal 15 for various materials that are saved and can be selected in the device The gravimetric moisture H is indicated as a percentage on the y axis and the associated radar transit time...

Page 18: ...el shift offset in the PLC The parameter has different names depending on the PLC e g initial load zero point offset measuring range etc Perform a parallel shift offset in the PLC Contact the PLC manu...

Page 19: ...se grained or hydrophobic materials water can escape directly onto the measurement surface and therefore cause a high moisture value Enter limit values in the PLC Contact the PLC manufacturer Incorrec...

Page 20: ...www addresses endress com 71550111 71550111...

Reviews: