Electrical connection
Hauser
19
6.3
Electrical connection - example, sensor cable with
10-pin socket (sensor side) and ferrules at the end of
the cable
6 7 8
5
2
1
4
3
2
1
2
1
3
4
3 4
A
B
C
1 2
A0037418
14
Connection example
A
Sensor
B
PLC / distribution box
C
Remote display (optional)
1
0 V
DC
power supply
Wire color: blue (BU)
2
12 to 24 V
DC
stabilized power supply
Wire color: red (RD)
3
IMP-Bus RT
Wire color: gray (GY) / pink (PK)
4
IMP-Bus COM
Wire color: blue (BU) / red (RD)
5
1st current output (+), analog
Wire color: green (GN)
6
1st current output (-), analog
Wire color: yellow (YE)
7
2nd current output (+), analog
Wire color: pink (PK)
8
2nd current output (-), analog
Wire color: gray (GY)
The moisture content determined and the conductivity / temperature can either be
fed directly into a PLC via analog outputs 0 to 20 mA/4 to 20 mA or queried via the
serial interface (IMP-Bus).
6.4
Post-connection check
Is the device or cable undamaged (visual inspection)?
Does the supply voltage match the specifications on the nameplate?
Are the connections established correctly and protected against mechanical influences?