Solicap M FTI55, FTI56
Installation
Hauser
19
3.7
Rod probes FTI55
!
Note!
Total length of the probe from the start of the thread: L = L1 + L3
(+ 125 mm with active buildup compensation)
Rod probe
partially insulated
Rod probe
fully insulated
Rod probe
with inactive length
partially/fully insulated
Rod probe with active
buildup compensation
partially insulated
TI418Fen39
H2
77
77
66
92
H3
25
25
25
25
Across flats (AF)
50
50
50
50
Total length (L)
200...4000
200...4000
300...6000
225...4000
Active rod length (L1)
200...4000
200...4000
200...4000
200...4000
Inactive rod length (L3)
-
-
200...2000
-
ø Inactive length
-
-
43
-
Length of partial insulation (L2)
75
-
75 / -
75
Probe rod diameter (with insulation)
18
(25)
18
(25)
18
(25)
18
(25)
ø Active build-up compensation/
length
- / -
- / -
- / -
36/
125
Lateral loading capacity (Nm) at 20 °C
300
300
300
200
Maximum process temperature (°C)
180
80
180/80
180
For use in mounting nozzles
-
-
X
-
In the event of condensate on tank
ceiling
-
-
X
X
X = recommended
L1/L2/L3
NPT / R
H2
L1
L2
L1
L1
L3
L2
L1
125
L2
H2
H3
H2
H3
H2
H2H2
H3
H2H2
H3
AF
AF
AF
AF
Length tolerance
Up to 1 m: 0 to –5 mm
> 1 m to 3 m: 0 to –10 mm
> 3 m to 6 m: 0 to –20 mm