Smartec CLD18
Commissioning
Hauser
27
Parameter
Possible settings
Description
Temp. unit
°C
°F
Setting for the temperature unit
Hold release
0 to 600 s
0 s
Prolongs the device hold when the hold condition no longer
applies
Alarm delay
0 to 600 s
0 s
Time delay after which an alarm is output
This suppresses alarm conditions that are present for a period
that is shorter than the alarm delay time.
Input
Setting for the inputs
Cell const.
Read only
Displays the cell constant
Inst. factor
0.1 to 5.0
1.0
The effects of the distance from the wall can be corrected
with the installation factor → 29
Unit
Auto
, μS/cm, mS/cm Unit of conductivity
"auto" automatically switches between μS/cm and mS/cm.
Damping main
0 ... 60 s
0 s
Setting for the damping
Temp. comp.
Off
,
Linear
Setting for temperature compensation
Alpha coeff.
1.0 to 20.0 %/K
2.1 %/K
Coefficient for linear temperature compensation
Ref. temp.
+10 to +50 °C
25 °C
Enter the reference temperature
Process check
The process check checks the measuring signal for stagnation. An
alarm is triggered if the measuring signal does not change over a
specific period (several measured values).
Function
On
,
Off
‣
Switch the process check on or off.
Duration
1 to 240 min
60 min
The measured value must change within this time as otherwise
an error message is triggered.
Observation width 1 to 20 %
0.0 %
Bandwidth for the process check
Analog output
Setting for analog outputs
Current range
4-20 mA
0-20 mA
Current range for analog output
Out1 0/4 mA
0 to 2000000 μS/cm
0 μS/cm
‣
Enter the measured value at which the min. current value
(0/4 mA) is present at the transmitter output.
Out1 20 mA
0 to 2000000 μS/cm
0 μS/cm
‣
Enter the measured value at which the max. current value
(20 mA) is present at the transmitter output.
Out2 0/4 mA
-50 to 250 °C
0.0 °C
‣
Enter the measured value at which the min. current value
(0/4 mA) is present at the transmitter output.
Out2 20 mA
-50 to 250 °C
100.0 °C
‣
Enter the measured value at which the max. current value
(20 mA) is present at the transmitter output.