Adjustment and settings
Prothermo NMT532
44
Hauser
Element interval
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART device configuration → Element interval (14743)
Conditions
Kind of interval: Equal
Description
Sets the interval of each element.
Input unit
Numerical value
Factory setting
0 mm
Additional information
Read access: Operator
Write access: Maintenance
Select element
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) → NMT device
configuration → Element configuration → Select element (14734)
Description
The element to be configured is manually selected.
Input unit
1-16
Factory setting
1
Additional information
Read access: Operator
Write access: Maintenance
Zero adjust
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) → NMT device
configuration → Element configuration → Zero adjust (14735)
Description
Adjusts the offset of the selected element.
Input unit
Numerical value
Factory setting
0 (None)
Additional information
Read access: Operator
Write access: Maintenance
Element temperature
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) → NMT device
configuration → Element configuration → Element temp (14737)
Description
Displays the element temperature.
Additional information
Read access: Operator
Write access: -