Prothermo NMT539
Operation
Hauser
29
5.3.1
Primary value: VH00-VH09
Code
Display
Details
VH00
Liquid Temp
(Average liquid
temperature)
Item type
Read only
Range
–200 to 240 °C (–328 to 464 °F)
Display of the average temperature of the liquid phase
The measured liquid levels that are required for calculating the
average temperature of the liquid phase are provided by
Micropilot FMR Series (via NRF590) or NMS5, NMS7 or NMS8x.
VH01
Gas Temp
(Average gas
temperature)
Item type
Read only
Range
–200 to 240 °C (–328 to 464 °F)
Displays the average temperature of the measured gas (vapor) phase
Gas phase measurements that are required for calculating the
average gas phase temperature are provided by Micropilot FMR
series (via NRF590) or NMS5, NMS7 or NMS8x.
VH02
Measured Distance
(Liquid level)
Item type
Read only
Range
0 to 99
999 mm
Displays the liquid level inside the tank as configured by a level gauge.
If a level gauge is not connected, a directly input liquid level can be
used as a device test.
VH07
Temperature 0
(Element 0
temperature)
Item type
Read only
Tolerance
–1.0 to 1.1 °C (30.2 to 33.8 °F)
Checks that the measuring temperature resistor temperature
conversion is executed correctly.
VH09
Temperature 17
(Element 17
temperature)
Item type
Read only
This temperature is used for checking when shipping from the factory.
5.3.2
Temperature measurement elements 1: VH10-VH19
Code
Display
Details
VH10-19
Temperature 1-10
(Temperature of
elements 1 to 10)
Item type
Read only
Range
–200 to 240 °C (–328 to 464 °F)
Displays individual temperature measurement elements.
5.3.3
Temperature measurement elements 2: VH20-VH29
Code
Display
Details
VH20-25
Temperature 11-16
(Temperature of
elements 11 to 16)
Item type
Read only
Range
–200 to 240 °C (–328 to 464 °F)
Displays individual temperature measurement elements.
VH26
Selec. Ave Method
(Average
temperature
calculation method)
Item type
Selection
Selection
Standard / Advanced
Selects the method of average temperature calculation.