21
Hauser
Prosonic T - FMU 130, 131, 230, 231, 232
V0H3 Applications
0:
Liquid
including automatic stirrer
suppression
1:
Rapid level change
The height changes quickly.
2:
Liquid / dome cover including
automatic stirrer suppression.
The instrument is mounted
under a dome cover.
The maximum initial echo
factor is entered as standard.
3.
Coarse bulk solids
(grain size from 0.16
²
/ 4 mm)
4.
Conveyor belt
Calibration
Reset
A reset causes most of the instrument settings to return to the factory settings. The
following parameters are not affected by a reset:
• all linearization parameters (V2H0 – V2H3)
• the Tag number (VAH0)
• the m/ft selection (V8H2)
Note V8H2 Units of Length
• Units of length remain unchanged after carrying out a reset.
• They may only be entered directly after a reset. If the units of length are changed at
a later date, then all subsequent entries must be repeated.
Display:
V0H0:
Level in %
V0H8:
Distance in meters/feet
V0H9:
Level in meters/feet
Caution!
All entries which follow (linearization, current output, fixed target suppression) must be
in the same units as those of the calibration.
#
VH
Entry
Text
1
V9H5
333
H
Reset instrument
2
V8H2
(0 to 1)
H
Units of length
0:
meter
1: feet
3
V0H1
E (m/ft)
H
Empty calibration
4
V0H2
F (m/ft)
H
Full calibration
5
V0H3
H
Application
Full
100%
Empty
0%