Operating menu
Proservo NMS83
170
Hauser
Empty
Navigation
Setup → Empty
Description
Distance from reference point to zero position (tank bottom or datum plate).
User entry
0 to 100
000 mm
Factory setting
Dependent on the device version
Additional information
Read access
Operator
Write access
Maintenance
The reference point is the reference line of the calibration window.
Tank reference height
Navigation
Setup → Tank ref height
Description
Defines the distance from the dipping reference point to the zero position (tank bottom or
datum plate).
User entry
0 to 100
000 mm
Factory setting
Dependent on the device version
Additional information
Read access
Operator
Write access
Maintenance
Tank level
Navigation
Setup → Tank level
Description
Shows the distance from the zero position (tank bottom or datum plate) to the product
surface.
Additional information
Read access
Operator
Write access
-