Installation
Proline t-mass T 150
18
Hauser
4. If a Tri-Clamp process connection is used, determine seal thickness (E).
NOTICE!
The pipe wall thickness (B), socket height (D) and sealing thickness (E)
may not exceed the permitted height.
B + D + E may not be greater than 77 mm (3.03 in).
5. If a conical coupling process connection with a self-centering sealing ring is used,
determine seal thickness (E).
NOTE!
The pipe wall thickness (B), socket height (D) and sealing thickness (E) may
not exceed the permitted height.
B + D + E may not be greater than 77 mm (3.03 in).
6. If an aseptic liner or a conical coupling process connection is used, set the seal thickness
(E) to equal zero and do not take it into consideration.
NOTE!
The pipe wall thickness (B) and socket height (D) may not exceed the
permitted height.
B + D may not be greater than 77 mm (3.03 in).
Determining the customized length
‣
For all nominal diameters: 8 + B + D + E
Installation conditions for nipples
90° (±3°)
D
90° (±3°)
A0011843
2
Installation conditions for mounting bosses and threadolets
D
= 20.0 mm ± 0.5 mm (0.79 in ± 0.02 in)
‣
In the case of weld-in couplings with PEEK clamping ferrules, remove the clamping
ferrules before you commence welding to avoid heat damage from the welding process.