Proline t-mass T 150
Installation
Hauser
17
Order code for
"Material of insertion pipe; sensor"
, option CD
"….. inch customized length,
0.8μm, mechanically polished"
and option CE
"….. inch customized length, 0.4μm, mechanically
polished"
NOTICE
When ordering the customized length, it is necessary to declare the sensor length with
the following decimal accuracies:
‣
SI units (mm):
With a minimum of 1 decimal place. Example: 43.3 mm
‣
US units (in):
With a minimum of 2 decimal places. Example: 17.05 in
‣
When ordering, a maximum of 3 decimal places can be declared.
NOTICE
Certain dimensions are required for determining the customized length.
‣
Pay attention to information in the dimension drawings.
Preconditions
G
F
D
E
A
X
B
A0022058
A
Fixed insertion depth 8 mm (0.31 in) ±2 mm (0.08 in)
B
Pipe wall thickness
D
Socket height
E
Seal thickness
X
Length
G
Socket internal diameter
1. Determine pipe wall thickness (B).
2. Measure socket height (D).
3. Observe the maximum socket height D.
NOTE!
The pipe wall thickness (B) and socket height (D) may not exceed the
permitted height.
B + D may not be greater than 77 mm (3.03 in).