XA01148D
Proline Prowirl 200
18
Hauser
Position
Basic specification,
position 1, 2:
Approval
Type of protection
used
for cable entry
Description
1
Cable entry for
output 1
BC, BK, IC, IK,TC
B3, I5
Ex d
Ex d/Ex tb
In the case of device versions with a plastic
transport sealing plug, this plug does not meet
the explosion protection requirements and must
be replaced during installation by a suitable
entry that meets the approval specifications.
2
Cable entry for
output 2
BC, BK, IC, IK,TC
B3, I5
Ex d
Ex d/Ex tb
In the case of device versions with a metal
sealing plug, this plug is part of the device
approval and meets the requirements of the
explosion protection indicated on the
nameplate.
3
Cable entry of
the remote
display and
operating
module
FHX50
BC, BK, IC, IK,TC
B3, I5
Ex ia
Ex ia/Ex tb
The following applies for devices with basic
specification, position 1, 2 (Approval) = B3, I5:
In the case of device versions with a metal
sealing plug, this plug is part of the device
approval and meets the requirements of the
explosion protection indicated on the
nameplate.
In the case of device versions with a cable entry,
this entry has a separate component approval
and meets the requirements of the explosion
protection indicated on the nameplate.
Position
Description
4
Pressure compensation plug
L
WARNING
Housing degree of protection voided due to insufficient sealing of the
housing.
‣
Do not open - not a cable entry.
Potential equalization
NOTICE
Terminal for connection to potential equalization.
‣
Pay attention to the grounding concept of the facility.
1)
The labeling changes in connection with "Display; Operation" = "L" or "M": Ex tb[ia Da] IIIC Txx °C Db.