Proline Promass F 100 Modbus RS485
Technical data
115
Heartbeat Technology
Package
Description
Heartbeat Verification
+Monitoring
Heartbeat Verification
Meets the requirement for traceable verification to DIN ISO 9001:2008 Chapter
7.6 a) "Control of monitoring and measuring equipment".
• Functional testing in the installed state without interrupting the process.
• Traceable verification results on request, including a report.
• Simple testing process via local operation or other operating interfaces.
• Clear measuring point assessment (pass/fail) with high test coverage within the
framework of manufacturer specifications.
• Extension of calibration intervals according to operator's risk assessment.
Heartbeat Monitoring
Continuously supplies data, which are characteristic of the measuring principle, to
an external condition monitoring system for the purpose of preventive
maintenance or process analysis. These data enable the operator to:
• Draw conclusions - using these data and other information - about the impact
process influences (such as corrosion, abrasion, buildup etc.) have on the
measuring performance over time.
• Schedule servicing in time.
• Monitor the process or product quality, e.g. gas pockets.
Concentration
Package
Description
Concentration
measurement and special
density
Calculation and outputting of fluid concentrations
Many applications use density as a key measured value for monitoring quality or
controlling processes. The device measures the density of the fluid as standard and
makes this value available to the control system.
The "Special Density" application package offers high-precision density
measurement over a wide density and temperature range particularly for
applications subject to varying process conditions.
The observed density is converted to the concentration of a substance of a binary
mixture using the "Concentration Measurement" application package:
• Choice of predefined fluids (e.g. various sugar solutions, acids, alkalis, salts,
ethanol etc.)
• Common or user-defined units (°Brix, °Plato, % mass, % volume, mol/l etc.) for
standard applications.
• Concentration calculation from user-defined tables.
The measured values are output via the digital and analog outputs of the device.
16.14 Accessories
Overview of accessories available for order → 92
16.15 Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The
W@M Device Viewer
: Enter the serial number from the nameplate
)
• The
Hauser Operations App
: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.