Diagnostics and troubleshooting
Proline Promass E TB2 HART
88
Hauser
Diagnostics event
Event text
Remedial measures
Event level from the
factory
S912
Inhomogeneous
Fluid is inhomogeneous, e.g. gas or solid
content!
1. Check process conditions.
2. Increase system pressure.
In particular with out gassing fluids and/
or increased gas content,the following
measures are recommended to increase
system pressure:
• Install the instrument at the outlet side
of a pump.
• Install the instrument at the lowest
point of an ascending pipeline.
• Install a flow restriction, e.g. reducer
or orifice plate, downstream from the
instrument.
Warning
S913
Inhomogeneous
Oscillation amplitude limit.
The fluid properties do not allow a precise
measurement.
Cause: Process fluid is very inhomogeneous (gas
or solid content)
1. Check process conditions
2. Increase supply
3. Check main electronic or sensor.
Warning
1)
Event level can be changed: Chapter 12.5.1 "Adapting event level"
12.6
Event logbook
12.6.1 Event history
A chronological overview of the event messages that have occurred is provided in the
Events list
submenu.
Navigation path
Diagnostics
®
Event logbook
®
Events list
F
I1091 Config. change
I1157 Mem.err. ev.list
F311 Electr. failure
/ ../Eventlist
0d01h19m10s
A0014008-EN
A maximum of 20 event messages can be displayed in chronological order. If the advanced
HistoROM function is enabled in the device (order option), up to 100 entries can be displayed.
The event history includes entries for:
• Diagnostics events (
In addition to the operation time of its occurrence, each event is also assigned a symbol that indicates
whether the event has occurred or is ended:
• Diagnostics event
–
: Event has occurred
–
: Event has ended
• Information event
: Event has occurred