Operating Instructions

Proline Promass E TB2

HART

Coriolis mass flow measuring system

BA01027D/06/EN/06.10
71112405

Valid as of version
01.00.zz (Device software)

Summary of Contents for Proline Promass E TB2 HART

Page 1: ...Operating Instructions Proline Promass E TB2 HART Coriolis mass flow measuring system BA01027D 06 EN 06 10 71112405 Valid as of version 01 00 zz Device software...

Page 2: ...s or the facility read the Basic safety instructions section carefully as well as all other safety instructions in the document that are specific to working procedures The manufacturer reserves the ri...

Page 3: ...6 2 4 Turning the transmitter housing 22 6 2 5 Turning the display module 23 6 3 Post mounting check 24 7 Electrical connection 25 7 1 Preparing the measuring device for the electrical connection 25 7...

Page 4: ...12 1 General troubleshooting 81 12 2 Diagnostics event on local display 82 12 2 1 Diagnostic message 82 12 2 2 Calling up remedial measures 83 12 3 Diagnostics event in the operating tool 84 12 4 Diag...

Page 5: ...Version for experts Reference for your parameters The document provides a detailed explanation of each individual parameter in the 2nd part of the operating menu the Expert menu It contains all the d...

Page 6: ...angerous situation Failure to avoid this situation can result in minor or medium injury NOTICE A0011192 EN NOTE This symbol contains information on procedures and other facts which do not result in pe...

Page 7: ...ions that are forbidden A0011193 Tip Indicates additional information A0011194 Reference to documentation Refers to the corresponding device documentation A0011195 Reference to page Refers to the corr...

Page 8: ...ed in the Operating Instructions and supplementary documentation the measuring device may be used for the following measurements only Measured process variables mass flow density temperature Calculate...

Page 9: ...e not permitted and can lead to unforeseeable dangers If despite this modifications are required consult with Endress Hauser Repair To ensure continued operational safety and reliability Carry out rep...

Page 10: ...nics module 7 Terminals pluggable spring terminals 8 Connection compartment cover 9 Sensor 3 2 Registered trademarks HART Registered trademark of the HART Communication Foundation Austin USA KALREZ VI...

Page 11: ...Proline Promass E TB2 HART Product description Endress Hauser 11 3 3 Patents US patents EP patents 5 610 342 0 701 213 6 513 393 1 055 102 6 412 35 1 054 240 6 651 513 6 412 355 1 054 240 7 555 397...

Page 12: ...e on the delivery note 1 identical to the order code on the product sticker 2 A0013695 A0013696 A0013698 Are the goods undamaged A0013696 i i DELIVERY NOTE A0013699 Do the nameplate data match the ord...

Page 13: ...ww endress com deviceviewer 4 2 1 Nameplates Transmitter Order code Ext ord cd Ser no Date i i Patents 322540 0001 1 2 3 4 5 6 7 8 9 10 11 12 14 15 16 17 13 A0013906 2 Example of a transmitter namepla...

Page 14: ...DG 3 A 14 Manufacturing date year month 15 Flow direction 16 2 D matrix code 17 Degree of protection e g IP NEMA 18 Approval information for explosion protection and Pressure Equipment Directive 19 Pe...

Page 15: ...igher than the suspension points of the webbing slings Risk of injury if the measuring device slips Secure the measuring device from rotating or slipping Observe the weight specified on the packaging...

Page 16: ...95 EC RoHS Packaging Wood crate treated in accordance with ISPM 15 standard which is confirmed by the affixed IPPC logo or Carton in accordance with European Packaging Directive 94 62EC recyclability...

Page 17: ...ising from accumulation of gas bubbles in the measuring tube avoid the following mounting locations in the pipe Highest point of a pipeline Directly upstream of a free pipe outlet in a down pipe Insta...

Page 18: ...iction mm in mm in 8 3 8 6 0 24 15 10 0 40 25 1 14 0 55 40 1 22 0 87 50 2 28 1 10 Orientation The direction of the arrow on the sensor nameplate helps to match the sensor with the direction of flow di...

Page 19: ...precautions need to be taken for fittings which create turbulence such as valves elbows or T pieces as long as no cavitation occurs 19 Installation dimensions For the dimensions and installation leng...

Page 20: ...he heat convection from the sensor to the transmitter A wide range of materials can be used to provide the required thermal insulation NOTICE Electronics overheating on account of thermal insulation O...

Page 21: ...ess d 0 35 mm d 0 014 in Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations 6 1 3 Additio...

Page 22: ...k 4 Remove stick on label on the electronics compartment cover 6 2 3 Mounting the measuring device WARNING Danger due to insufficient process sealing Ensure that the inside diameter of the seals is gr...

Page 23: ...tment from the transmitter housing 3 Optional pull out the display module with a gentle rotation movement 4 Rotate the display module into the desired position Max 8 45 in each direction 5 Without dis...

Page 24: ...perature range 19 Measuring range 97 m Has the correct orientation for the sensor been selected 18 According to sensor type According to fluid temperature According to fluid properties outgassing with...

Page 25: ...able diameter Included cable glands M20 1 5 with cable 6 to 12 mm 0 24 to 0 47 in Plug in spring terminals for wire cross sections 0 5 to 2 5 mm2 20 to 14 AWG 7 1 3 Preparing the measuring device 1 Re...

Page 26: ...0 mm 0 4 in For stranded cables also attach wire end ferrules 6 Firmly tighten the cable glands 7 2 1 4 3 A0013837 open 1 2 2 wire Flow Spare part 71108602 3 4 HART 4 20 mA 4 20 mA 04 05 HART 4 20 mA...

Page 27: ...ments for the supply unit Supply voltage Order version output US supply voltage Imax output current 8E2B C 18 to 30 V DC 22 0 mA Supply voltage range without load For information on the supply voltage...

Page 28: ...neatly and correctly Dry clean or replace the seals if necessary 2 Tighten all housing screws and screw covers 3 Firmly tighten the cable glands 4 So that any moisture that occurs does not reach the...

Page 29: ...connection Endress Hauser 29 m If supply voltage is present is the device ready for operation and do values appear on the display module m Are all housing covers installed and firmly tightened m Is t...

Page 30: ...ration options 8 1 Overview of operation options E 1 2 3 4 5 A0013902 1 Local operation via display module 2 Computer with operating tool e g FieldCare AMS Device Manager SIMATIC PDM 3 Field Xpert SFX...

Page 31: ...king status Access status display Output n Operating menu for experts Language Display operat Access status display Locking status Operation Setup Wizard 1 Wizard 2 Wizard n Advanced setup Enter acces...

Page 32: ...ntly pending diagnostic messages Event logbook submenu Contains up to 20 or 100 order option event messages that have occurred Device information submenu Contains information for identifying the devic...

Page 33: ...ion check The device is in service mode e g during a simulation A0013958 Out of specification The device is operated Outside of its technical specifications e g during startup or a cleaning Outside of...

Page 34: ...le A0013949 Channel 2 of a process variable A0013951 Channel 4 of a process variable Status symbols symbols for event level The status symbol pertains to a diagnostics event that is relevant to the di...

Page 35: ...nu In the wizard Display symbol for wizard Omission symbol for operating menu levels in between Name of current Submenu Wizard Parameters Examples A0013973 Display A0013968 Display For more informatio...

Page 36: ...rd A0013972 Parameters within a wizard No display symbol exists for parameters in submenus Lock symbols A0013963 Parameter locked When displayed in front of a parameter name indicates that the paramet...

Page 37: ...ess keys simultaneously Exits the wizard and takes you to the next higher level General operating functions A0013953 Minus Enter key combination press and hold down the keys simultaneously Reduces the...

Page 38: ...8 Deletes one character immediately to the left of the input position Operating elements General operating functions A0013969 Minus key In the input mask moves the selection bar to the left backwards...

Page 39: ...Calling up and closing the context menu The user is in the measured value display 1 Press F for 2 s The context menu opens XXXXXXXXXX kg h 20 50 Setup Conf backup disp Simulation A0014003 EN 2 Press...

Page 40: ...w with symbols and operating elements 35 Example Setting the number of displayed measured values to 2 values XXXX XXX X X 20 50 XXXX XXX X X 2 s 0104 1 Language 0091 1 0098 1 0098 1 0098 1 Language Di...

Page 41: ...of the parameter and thus support fast and reliable commissioning Calling up and closing the help text The user is in the navigation view and the selection bar is on a parameter 1 Press F for 2 s The...

Page 42: ...er 20 mA value to 20 kg s 0 1 2 3 5 6 7 8 9 4 20 kg s 0 1 2 3 5 6 7 8 9 4 100 000 0 1 2 3 5 6 7 8 9 4 100 000 0 1 2 3 5 6 7 8 9 4 0 1 2 3 5 6 7 8 9 4 0 1 2 3 5 6 7 8 9 4 2 0 1 2 3 5 6 7 8 9 4 2 kg s k...

Page 43: ...n incorrect access code is entered the user obtains the access rights of the Operator role The user role with which the user is currently logged on is indicated by the Access status display parameter...

Page 44: ...ng the S O F keys After enabling the keypad lock XXXXXXXXXX kg h 20 50 Keylock on A0014000 EN After disabling the keypad lock XXXXXXXXXX kg h 20 50 Keylock off A0014001 EN If the user attempts to acce...

Page 45: ...iles See data 49 8 4 2 FieldCare Function scope FDT based plant asset management tool from Endress Hauser It can configure all smart field units in a system and help you manage them By using the statu...

Page 46: ...e data 49 8 4 4 SIMATIC PDM Function scope SIMATIC PDM is a standardized manufacturer independent program from Siemens for the operation configuration maintenance and diagnosis of intelligent field de...

Page 47: ...via HART protocol 1 PLC programmable logic controller 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubox FXA191 FXA195 and Field Communicator 375 475 4 Fi...

Page 48: ...Proline Promass E TB2 HART 48 Endress Hauser Via CDI interface 1 2 A0014019 10 Operating options via CDI interface 1 Computer with operating tool FieldCare 2 Commubox FXA291 connected to the CDI inter...

Page 49: ...description file for each individual operating tool with source Operating tools Operating tool Sources for obtaining device descriptions Field Xpert SFX100 Use update function of handheld terminal Fi...

Page 50: ...w Volume flow Corrected volume flow Density Reference density Temperature Measured values for SV TV QV secondary tertiary and quarternary variable Mass flow Volume flow Corrected volume flow Density R...

Page 51: ...vice After a successful startup the local display switches automatically from the startup display to the measured value display If nothing appears on the local display or a diagnostic message is displ...

Page 52: ...vigation to the Setup menu XXXX XXX X X 20 50 Language 0104 1 Display operat Setup Diagnostic Display operat Setup Main menu English Main menu 1 2 3 4 Select medium Setup Curr output 2 Curr output 1 A...

Page 53: ...the wizard Reference sound velocity Pressure compensation Temperature coefficient sound velocity Select medium Liquid Gas Select gas type Air Others Off Fixed Pressure value End of wizard A0013795 EN...

Page 54: ...99999 m s 0 m s Temperature coefficient sound velocity Enter the temperature coefficient of the sound velocity of the gas Max 15 digit positive floating point number 0 m s K Pressure compensation Enab...

Page 55: ...nu Current output 1 Setup menu Current output 2 Sequence of the wizard Part 1 corresponding part 2 next page Assign current output Off Mass flow Volume flow Density Temperature Mass flow unit Volume f...

Page 56: ...t Fixed reference density Corrected volume flow Reference density Linear expansion coefficient Square expansion coefficient Temperature unit Reference temperature Reference density unit 20 mA value Fa...

Page 57: ...the unit for density Result The selected unit applies for Current outputs Low value partial filled pipe detection High value partial filled pipe detection Simulation process variable Density adjustme...

Page 58: ...ariable Unit picklist Country dependent C Celsius F Fahrenheit Reference temperature Enter the reference temperature for calculating the reference density Max 15 digit floating point number with sign...

Page 59: ...splay Value 1 display 0 bargraph value 1 100 bargraph value 1 Value 2 display Value 3 display 0 bargraph value 3 100 bargraph value 3 Value 4 display End of wizard A0013797 EN 14 Display wizard in the...

Page 60: ...h sign Depends on country and nominal diameter Value 2 display Select the measured value shown on the local display None Mass flow Volume flow Corrected volume flow Density Reference density Temperatu...

Page 61: ...ne Mass flow Volume flow Corrected volume flow Density Reference density Temperature Totalizer 1 Totalizer 2 Totalizer 3 Current output 1 Current output 2 None 10 4 4 Configuring the output conditioni...

Page 62: ...s variable Pressure shock suppression On value low flow cutoff Off value low flow cutoff Operating mode totalizer 1 Slight Moderate Strong Display damping Display damping Display damping Damping outpu...

Page 63: ...fluctuations in the measured value caused by process conditions 0 to 999 s 1 s Measuring mode output 1 2 Select the measuring mode for current output Forward flow Forward reverse flow Reverse flow com...

Page 64: ...A0013799 EN 16 Low flow cut off wizard in the Setup menu Parameter overview with brief description Parameters Description Selection User entry Factory setting Assign process variable Select the proce...

Page 65: ...eter overview with brief description Parameters Description Selection User entry Factory setting Assign process variable Select a process variable to detect empty or partially filled pipes Off Density...

Page 66: ...Display operat Setup Diagnostic Display operat Setup Main menu English Main menu 1 2 3 4 Select medium Setup Curr output 2 Curr output 1 Low flow cut off Setup Advanced setup Partial pipe det 5 Ent a...

Page 67: ...or special characters e g In the factory configuration the tag name appears at the top left in the header 1 XXXXXXXXX A0013375 1 Device tag The number of characters displayed depends on the character...

Page 68: ...riable Unit picklist Country dependent l h gal min us Volume Select the unit for volume Result The selected unit is taken from Volume flow unit Country dependent l gal us Corrected volume flow unit Se...

Page 69: ...us F Fahrenheit Length unit Select the unit of length for the nominal diameter Unit picklist Country dependent mm in Pressure unit Select the unit for pipe pressure Unit picklist Country dependent bar...

Page 70: ...es the picklist of the Unit parameter Off Mass flow Volume flow Corrected volume flow Mass flow Unit One of the following options is selected in the Assign process variable parameter Mass flow Volume...

Page 71: ...0 bargraph value 1 59 Decimal places 1 Value 2 display 59 Decimal places 2 Value 3 display 59 0 bargraph value 3 59 100 bargraph value 3 59 Decimal places 3 Value 4 display 59 Decimal places 4 Display...

Page 72: ...special characters Separator Select the decimal separator for numerical values point comma point 10 6 Configuration management After commissioning you can save the current device configuration copy it...

Page 73: ...ation device alarm Parameter overview with brief description Parameters Prerequisite Description Selection User entry Factory setting Assign simulation process variable Select a process variable for t...

Page 74: ...e symbol appears in front of all write protected parameters Parameters that can always be changed The write protection does not include certain parameters that do not affect the measurement Despite th...

Page 75: ...sening the securing clamp and unscrewing the electronics compartment cover 2 Pull out the display module with a gentle rotation movement 3 A0013909 To make it easier to access the lock switch attach t...

Page 76: ...f the measured value display and in the navigation view in front of the parameters 5 Feed the spiral cable into the gap between the housing and main electronics module and plug the display module into...

Page 77: ...lues You can read all measured values using the Measured value menu Navigation path Diagnostics Measured val The following three submenus are available Process variable submenu Process variab Mass flo...

Page 78: ...et all totalizers Navigation path Display operat menu Operation Operation submenu Operation Control totalizer 1 Preset value 1 Control totalizer 2 Preset value 2 Control totalizer 3 Preset value 3 Res...

Page 79: ...ta logging submenu Data logging Assign channel 1 Assign channel 2 Assign channel 3 Assign channel 4 Logging interval Clear logging data Display channel 1 Display channel 2 Display channel 3 Display ch...

Page 80: ...ration Proline Promass E TB2 HART 80 Endress Hauser If the length of the logging interval or the assignment of the process variables to the channels is changed the content of the data logging is delet...

Page 81: ...e Order spare part 91 I O electronics module is defective Order spare part 91 Device shows correct value on local display but signal output is incorrect though in the valid current range Configuration...

Page 82: ...ng elements Status signals A0013956 Failure A device error is present The measured value is no longer valid A0013957 Maintenance required Maintenance is required The measured value is still valid A001...

Page 83: ...ional pending diagnostic messages can be shown in the Diagnose list submenu 84 Past diagnostic messages that are no longer pending are shown in the Event logbook submenu 88 Operating elements Operatin...

Page 84: ...07 Failure F Maintenance required M Function check C Out of specification S A0014014 EN Calling up remedial measures 1 Navigate to the Diagnostics menu In the Actual diagnostics parameter the diagnost...

Page 85: ...ert menu System Diagn handling Event level Event no xxx 0723 1 Event no 044 Event level Event no 801 Event no 274 Warning A0014048 EN You can assign the following options to the diagnostics event as t...

Page 86: ...les incompatible 1 Check electronic modules 2 Change I O or main electronics module Alarm 1 F261 Electr modules 1 Restart device 2 Check electronic modules 3 Change I O Modul or main electronics Alarm...

Page 87: ...ulation Warning 12 5 5 Diagnostics for the process Diagnostics event Event text Remedial measures Event level from the factory S801 Supply voltage Supply voltage too low Increase voltage Warning 1 M80...

Page 88: ...gas or solid content 1 Check process conditions 2 Increase supply 3 Check main electronic or sensor Warning 1 1 Event level can be changed Chapter 12 5 1 Adapting event level 12 6 Event logbook 12 6 1...

Page 89: ...intenance required M Function check C Out of specification S Information I 12 6 3 Overview of information events Unlike a diagnostic event an information event is displayed in the event logbook only a...

Page 90: ...hooting Proline Promass E TB2 HART 90 Endress Hauser Information event Event text I1221 Error during zero point adjustment I1222 Zero point adjustment OK I1227 Sensor emergency mode enabled I1228 Sens...

Page 91: ...umentation XA and certificates Document every repair and each conversion and enter them into the W M life cycle management database 13 2 Spare parts A few interchangeable measuring device components a...

Page 92: ...Repair Proline Promass E TB2 HART 92 Endress Hauser 13 3 Endress Hauser services For information about service and spare parts contact your Endress Hauser distributor...

Page 93: ...rve the maximum permitted fluid temperature for the measuring device 103 14 2 Measuring and test equipment Endress Hauser offers a wide variety of measuring and test equipment such as W M or device te...

Page 94: ...subsequently from Endress Hauser are available for the device Detailed information on the order code in question is available from your local Endress Hauser sales center or on the product page of the...

Page 95: ...zardous to health e g flammable toxic caustic or carcinogenic Costs incurred for waste disposal and injury due to inadequate cleaning will be charged to the owner operator Complete all fields of the D...

Page 96: ...device and Connecting the measuring device in the logically reverse sequence Observe the safety instructions 17 2 Disposing of the measuring device WARNING Danger to personnel and environment from flu...

Page 97: ...ple Measuring system The device consists of a transmitter and a sensor One device version is available Compact version transmitter and sensor form a mechanical unit For information on the structure of...

Page 98: ...03 Operable flow range Greater than 1000 1 Flow rates above the preset full scale value do not overload the amplifier so the totalized values are registered correctly 18 4 Output Output signal 2 wire...

Page 99: ...o 660 mW Operation with current output 1 and 2 130 to 1320 mW Power supply failure Totalizers stop at the last value measured Configuration is retained in the HistoROM Error messages incl total operat...

Page 100: ...easured error In addition to the values indicated the measured error at the current output is typically 4 mA o r of reading 1 g cm3 1 kg l T fluid temperature Mass flow and volume flow liquids 0 25 o...

Page 101: ...0 1 72 2 646 3 100 1 180 6 615 1 2 25 1 720 26 46 0 3 10 1 1800 66 15 0 25 2 1 9000 330 75 0 25 Repeatability o r of reading 1 g cm3 1 kg l T fluid temperature Mass flow and volume flow liquids 0 125...

Page 102: ...between calibration pressure and process pressure o r of reading DN o r bar mm in 8 3 8 no influence 15 no influence 25 1 no influence 40 1 no influence 50 2 0 009 Design fundamentals Dependent on the...

Page 103: ...nsor 40 to 140 C 40 to 284 F Seals No internal seals Medium density 0 1 to 2000 kg m3 0 2 to 4410 lb m3 Medium pressure range nominal pressure An overview of the material load diagrams pressure temper...

Page 104: ...sizing tool 94 Pressure loss in SI units Pressure loss depends on the fluid properties and on the flow rate The following formulas can be used to approximately calculate the pressure loss Reynolds nu...

Page 105: ...the device see the Technical Information document Mechanical construction section Rupture disk optional Rupture disks cannot be combined with the separately available heating jacket 94 Weight Weight i...

Page 106: ...nless steel 1 4404 316L For all process connections except flanges as per JIS B2220 Stainless steel SUS 316L For flanges as per JIS B2220 List of all available process connections 107 Measuring tubes...

Page 107: ...11864 1 Form A DIN 11864 2 Form A flat flange with groove For information on the materials of the process connections 106 18 11 Human interface Local display SD02 Display elements 4 line display Form...

Page 108: ...parate Safety Instructions XA document Reference is made to this document on the nameplate Hygienic compatibility 3A approval Other standards and guidelines EN 60529 Degrees of protection by housing I...

Page 109: ...pecified in Annex I of the Pressure Equipment Directive 97 23 EC Devices not bearing this marking PED are designed and manufactured according to good engineering practice They meet the requirements of...

Page 110: ...e measuring device 52 Communication specific data 49 Connecting cable Requirements 25 Connecting the measuring device 25 Connection see Electrical connection Connection preparations 25 Connection tool...

Page 111: ...elease date 49 Version 49 Flow direction 18 22 Flow limit 103 Fluids 8 Function check 51 Function scope AMS Device Manager 46 Field Communicator 46 Field Communicator 475 46 Field Xpert 45 FieldCare 4...

Page 112: ...g device configuration 52 For specific settings 66 Mounting 17 Mounting dimensions see Installation dimensions Mounting location 17 Mounting preparations 22 Mounting requirements Inlet and outlet runs...

Page 113: ...instructions 8 Safety symbols 6 Safety icons see Safety Seals Medium temperature range 103 Secondary containment pressure range 103 Sensor Medium temperature range 103 Mounting 22 Sensor heating 20 Se...

Page 114: ...Terminals 99 Text editor 37 Thermal insulation 20 Tool tip see Help text Tools For electrical connection 25 For mounting 22 Transport 15 Transmitter Connecting the signal cables 26 Turning the displa...

Page 115: ..._____________________________________________________________________________________ We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge We fu...

Page 116: ...www endress com worldwide 71112405 71112405 BA01027D 06 EN 06 10 71112405 EH COSIMA ProMoDo...