Brief Operating Instructions

Proline Promass A 100

Coriolis flowmeter

These Instructions are Brief Operating Instructions; they do not

replace the Operating Instructions included in the scope of

supply.
For detailed information, refer to the Operating Instructions

and other documentation on the CD-ROM provided or visit

"www.endress.com/deviceviewer".

KA01144D/06/EN/02.13

71231777

Summary of Contents for Proline Promass A 100

Page 1: ...e Instructions are Brief Operating Instructions they do not replace the Operating Instructions included in the scope of supply For detailed information refer to the Operating Instructions and other do...

Page 2: ...2 Mounting the measuring device 21 6 3 Post installation check 22 7 Electrical connection 23 7 1 Connection conditions 23 7 2 Connecting the measuring device 31 7 3 Hardware settings 33 7 4 Ensuring t...

Page 3: ...rical symbols Symbol Meaning A0011197 Direct current A terminal to which DC voltage is applied or through which direct current flows A0011198 Alternating current A terminal to which alternating voltag...

Page 4: ...at are preferred A0011184 Forbidden Indicates procedures processes or actions that are forbidden A0011193 Tip Indicates additional information A0011194 Reference to documentation Refers to the corresp...

Page 5: ...2 Designated use Application and media The measuring device described in these Instructions is intended only for flow measurement of liquids and gases Depending on the version ordered the measuring d...

Page 6: ...te changes in the temperature concentration or level of contamination in the process can alter the corrosion resistance properties Residual risks The external surface temperature of the housing can in...

Page 7: ...ctice to meet state of the art safety requirements has been tested and left the factory in a condition in which it is safe to operate It meets general safety standards and legal requirements It also c...

Page 8: ...Modbus RS485 and PROFIBUS DP communication types 2 3 4 1 A0017609 1 Important components of a measuring device 1 Transmitter housing cover 2 Main electronics module 3 Transmitter housing 4 Sensor In t...

Page 9: ...ser 9 3 1 2 Device version with EtherNet IP communication type 2 3 4 5 1 A0017257 2 Important components of a measuring device 1 Transmitter housing cover 2 I O electronics module for EtherNet IP 3 Ma...

Page 10: ...ance and product identification Proline Promass A 100 10 Endress Hauser 4 Incoming acceptance and product identification 4 1 Incoming acceptance A0015502 1 2 1 2 A0013843 A0013695 A0015502 A0013698 A0...

Page 11: ...numbers from nameplates in W M Device Viewer www endress com deviceviewer All information about the measuring device is displayed For an overview of the scope of the Technical Documentation provided...

Page 12: ...of the measuring device is higher than the suspension points of the webbing slings Risk of injury if the measuring device slips Secure the measuring device from rotating or slipping Observe the weigh...

Page 13: ...Proline Promass A 100 Storage and transport Endress Hauser 13 A0015604 A0015605...

Page 14: ...by the construction of the device 6 1 1 Mounting position Mounting location A0015595 Installation in down pipes However the following installation suggestion allows for installation in an open vertic...

Page 15: ...user 15 1 2 3 4 5 A0015596 4 Installation in a down pipe e g for batching applications 1 Supply tank 2 Sensor 3 Orifice plate pipe restriction 4 Valve 5 Batching tank DN orifice plate pipe restriction...

Page 16: ...low process temperatures may reduce the ambient temperature To maintain the minimum ambient temperature for the transmitter this orientation is recommended 2 Applications with high process temperatur...

Page 17: ...pipe Downstream from pumps no danger of vacuum A0015594 Thermal insulation In the case of some fluids it is important that the heat radiated from the sensor to the transmitter is kept to a minimum A...

Page 18: ...ency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations 6 1 3 Special mounting instructions Rupture disk Make sure that the functio...

Page 19: ...tailed information about using a rupture disk refer to the Operating Instructions for the device on the CD ROM provided Wall mounting WARNING Incorrect sensor mounting Risk of injury if measuring tube...

Page 20: ...ass A 100 20 Endress Hauser A0019631 Horizontal Device standing on a solid support base A0019632 Post retainer The post retainer mounting kit is used to secure the device to a pipe or post order code...

Page 21: ...ro point obtained in this way is printed on the nameplate of the measuring device Calibration takes place under reference conditions Therefore a zero point adjustment in the field is generally not req...

Page 22: ...transmitter housing so that the cable entries do not point upwards A0013964 6 3 Post installation check Is the device undamaged visual inspection Does the measuring device conform to the measuring poi...

Page 23: ...pplicable federal national regulations Permitted temperature range 40 C 40 F to 80 C 176 F Minimum requirement cable temperature range ambient temperature 20 K Power supply cable Standard installation...

Page 24: ...section Shielding Copper braided shielding or braided shielding with foil shield When grounding the cable shield observe the grounding concept of the plant EtherNet IP The standard ANSI TIA EIA 568 B...

Page 25: ...h mm2 AWG m ft 1 5 16 200 656 2 5 14 300 984 Cable diameter Cable glands supplied M20 1 5 with cable 6 to 12 mm 0 24 to 0 47 in Spring terminals Wire cross sections 0 5 to 2 5 mm2 20 to 14 AWG With Sa...

Page 26: ...rier Promass 100 Power Communication Lift panel for bus termination 1 2 A A0016922 6 Safety Barrier Promass 100 with terminals 1 Non hazardous area and Zone 2 Div 2 2 Intrinsically safe area 7 1 4 Pin...

Page 27: ...and Zone 2 Div 2 on the device side 3 2 4 1 5 A0016811 Pin Assignment Coding Plug socket 1 B Socket 2 A PROFIBUS DP 3 4 B PROFIBUS DP 5 Grounding shielding Modbus RS485 Modbus RS485 intrinsically safe...

Page 28: ...s Ensure the cable shield is grounded to the potential matching line at multiple points Connect every local ground terminal to the potential matching line NOTICE In systems without potential matching...

Page 29: ...ures with regard to input wiring must be taken to allow unrestricted operation when EMC interference is present These measures have been taken into account for this device Operation in the event of di...

Page 30: ...ncy equalizing currents Damage to the bus cable shield Only ground the bus cable shield to either the local ground or the protective ground at one end Insulate the shield that is not connected 7 1 6 P...

Page 31: ...ting the transmitter The connection of the transmitter depends on the following order codes Housing version compact or ultracompact Connection version device plug or terminals 1 2 1 2 3 4 A B C A00169...

Page 32: ...Cable 2 Device plug for signal transmission 3 Device plug for supply voltage Connect the cable in accordance with the terminal assignment or the device plug pin assignment 26 7 2 2 Connecting the Safe...

Page 33: ...dous area 6 Non hazardous area and Zone 2 Div 2 7 Intrinsically safe area 8 Transmitter terminal assignment 7 3 Hardware settings 7 3 1 Setting the device address EtherNet IP The IP address of the mea...

Page 34: ...1 Set the desired IP address using the corresponding DIP switches on the I O electronics module Hardware addressing with the configured IP address is enabled after 10 s 2 Disable software addressing v...

Page 35: ...IP address 7 3 2 Enabling the terminating resistor PROFIBUS DP To avoid incorrect communication transmission caused by impedance mismatch terminate the PROFIBUS DP cable correctly at the start and en...

Page 36: ...risation Bus termination Bus polarisation Not used 4 1 2 3 A0021274 11 Termination using DIP switches on the I O electronics module for baud rates 1 5 MBaud Modbus RS485 To avoid incorrect communicati...

Page 37: ...auser 37 If the transmitter is used in the non hazardous area or Zone 2 Div 2 2 Not used 1 Write protection 4 Bus termination 3 Not used OFF ON 26 27 A B 220 W A0017610 12 Terminating resistor can be...

Page 38: ...all the requirements for the IP66 67 degree of protection Type 4X enclosure To guarantee IP66 67 degree of protection Type 4X enclosure carry out the following steps after the electrical connection 1...

Page 39: ...ned 31 Does the supply voltage match the specifications on the transmitter nameplate For device version with Modbus RS485 intrinsically safe does the supply voltage match the specifications on the nam...

Page 40: ...y The individual parts of the operating menu are assigned to certain user roles Each user role corresponds to typical tasks within the device lifecycle For detailed information about the operating phi...

Page 41: ...functional but simplified version of the operating menu structure starts in the Web browser 8 2 3 Establishing a connection Configuring the Internet protocol of the computer The following information...

Page 42: ...nter the default setting for the IP address 192 168 1 212 The login page appears Device tag Webserv language English Ent access code Access stat tool Maintenance 1 2 OK A0017362 1 Device tag 49 2 Pict...

Page 43: ...agnostic messages currently pending listed in order of priority Data management Data exchange between PC and measuring device Upload the configuration from the device XML format create configuration b...

Page 44: ...tool refer to the Operating Instructions for the device on the CD ROM provided 8 3 1 Via service interface CDI This communication interface is present in the following device version Order code for O...

Page 45: ...ccessing the integrated device Web server or with FieldCare operating tool with COM DTM CDI Communication TCP IP 3 Standard Ethernet connecting cable with RJ45 plug 1 2 3 A0021270 15 Connection for or...

Page 46: ...vice with access to the integrated Web server 2 Computer with Web browser e g Internet Explorer for accessing the integrated device Web server or with FieldCare operating tool with COM DTM CDI Communi...

Page 47: ...terface CDI RJ45 and FieldCare operating tool 1 Start FieldCare and launch the project 2 In the network Add a device The Add device window opens 3 Select the CDI Communication TCP IP option from the l...

Page 48: ...ructions BA00027S and BA00059S 9 System integration For system integration See Operating Instructions for the device on the CD ROM provided Applies only to device version with EtherNet IP communicatio...

Page 49: ...device address currently configured can be read out Operating Instructions for the device 10 4 Configuring the measuring device The Setup menu with its submenus is used for fast commissioning of the...

Page 50: ...specific access code access to the measuring device via the Web browser is protected as are the parameters for the measuring device configuration Navigation Setup menu Advanced setup Administration De...

Page 51: ...nal temperature Reference density All parameters for configuring the totalizer The parameter values are now read only and cannot be edited any more Via service interface CDI Via HART protocol Via Modb...

Page 52: ...915 For device version with PROFIBUS DP communication type OFF ON 1 2 3 4 5 6 7 8 9 10 1 2 4 8 16 32 64 Write protection Not used PROFIBUS address Software addressing A0021262 Setting the write protec...

Page 53: ...ion has been established and on the home page of the web browser once the user has logged on Remedial measures are provided for every diagnostic event to ensure that problems can be rectified quickly...

Page 54: ......

Page 55: ......

Page 56: ...www addresses endress com...

Reviews: