Proline Promass 80E, 83E
18
Hauser
Horizontal installation
When using a bent measuring tube and horizontal installation, the position of the sensor has to be
matched to the fluid properties!
a0004581
Horizontal installation for sensors with a bent measuring tube
1
Not suitable for fluids with entrained solids. Risk of solids accumulating.
2
Not suitable for outgassing fluids. Risk of air accumulating.
Zero point adjustment
All measuring devices are calibrated with state-of-the-art technology. The zero point determined in
this way is imprinted on the nameplate of the device. Calibration takes place under reference operating
conditions
ä
13. Consequently, the zero point adjustment is generally
not
necessary for Promass!
Experience shows that the zero point adjustment is advisable only in special cases:
• To achieve highest measuring accuracy also with very small flow rates.
• Under extreme process or operating conditions (e.g. very high process temperatures or very high
viscosity fluids).
Inlet and outlet runs
There are no installation requirements regarding inlet and outlet runs.
Length of connecting cable
Max. 20 meters (66 ft), remote version
Environment
Ambient temperature range
Sensor, transmitter:
• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)
!
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Storage temperature
–40 to +80 °C (–40 to +175 °F), pref20 °C (+68 °F)
Degree of protection
Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock resistance
According to IEC 68-2-31
Vibration resistance
Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR recommendation NE 21