Proline Promass 80E, 83E
16
Hauser
Installation
Mounting location
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
Therefore, avoid
the following mounting locations in the pipe installation:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipeline.
a0003605
Mounting location
Installation in down pipes
Notwithstanding the above, the installation proposal below permits installation in an open vertical
pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal
diameter prevent the sensor running empty while measurement is in progress.
a0003597
Installation in a down pipe (e.g. for batching applications)
1
Supply tank
2
Sensor
3
Orifice plate, pipe restriction (see Table following page)
4
Valve
5
Batching tank
1
2
3
4
5
DN
Orifice plate, pipe restriction
[mm]
[inch]
[mm]
[inch]
8
³
⁄
₈
"
6
0.24
15
½"
10
0.40
25
1"
14
0.55
40
1½"
22
0.87
50
2"
28
1.10
80
3"
50
2.00