5.4 Calling the Commissioning Quick Setup
All the functions needed for commissioning are called up automatically with the Quick Setup.
The functions can be changed and adapted to the process in question.
1.
F
→
Enter the function matrix (starting with measured value display)
2.
P
→
Select the group QUICK SETUP
F
→
Con
fi
rm selection
3.
QUICK SETUP COMMISSIONING function appears.
4.
Intermediate step if configuration is blocked:
P
→
Enter the code
80
(confirm with
F
) and thus enable configuration
5.
P
→
Go to Commissioning Quick Setup
6.
P
→
Select YES
F
→
Con
fi
rm selection
7.
F
→
Start Commissioning Quick Setup
8.
Configure the individual functions/settings:
– Via
P
-key, select option or enter number
– Via
F
-key, confirm entry and go to next function
– Via
Q
-key, return to Setup Commissioning function
(settings already made are retained)
Note! Observe the following when performing the Quick Setup:
• Configuration selection: Select the ACTUAL SETTING option
• Unit selection: This is not offered again for selection after configuring a unit
• Output selection: This is not offered again for selection after configuring an output
• Automatic configuration of the display: select YES
– Main line = Mass flow
– Additional line = Totalizer 1
• If asked whether additional Quick Setups should be executed: select NO
All the available functions of the measuring device and their configuration options as well as
additional Quick Setups, if available, are described in detail in the "Description of Device
Functions" Operating Instructions. The related Operating Instructions can be found on the
CD-ROM.
The measuring device is ready for operation on completion of the Quick Setup.