Process error messages
!
Note!
Also observe the information on
9.3.1
Displaying the device status on PROFIBUS
For more information, refer to
9.3.2
List of process error messages
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
Qu
a
lit
y
cod
e
(
H
E
X
)
M
e
as
ure
d
val
u
e
s
tatus
Q
u
al
it
y s
tatus
Q
u
al
it
y
s
ubstatus
Li
m
its
P = Process error
$
= Fault message (
with
and effect on operation)
! = Notice message (
without
an effect on operation)
401
P: EMPTY PIPE
$
: # 401
0x03
BAD
Non specific
(uncertain
status)
Constant
Empty Pipe
detected
Cause:
Measuring tube partially filled or empty.
Remedy:
1.
Check the process conditions of the plant.
2.
Fill the measuring tube.
461
P: ADJ. NOT OK
!: # 461
0x40
0x41
0x42
UNCERT
AIN
Non specific
(uncertain
status)
O.K.
Low
High
EPD adjustment
not possible
Cause:
EPD adjustment not possible because the fluid's
conductivity is either too low or too high.
Remedy:
The EPD function cannot be used with fluids of this
nature.
463
P: FULL = EMPTY
$
: # 463
0x40
0x41
0x42
UNCERT
AIN
Non specific
(uncertain
status)
O.K.
Low
High
EPD adjustment
wrong
Cause:
The EPD calibration values for empty pipe and full pipe
are identical and therefore incorrect.
Remedy:
Repeat adjustment, making sure procedure is correct.
467
S: AO-Block Error
$
: # 467
0x0F
BAD
Device
Failure
Constant
AO-Block Error
Cause:
The value imported cyclically using the AO block is
invalid (status = BAD).
Remedy:
Ensure that the status of the imported value = GOOD.
Process variables affected
– calculated mass flow
– target mass flow
– % target mass flow
– carrier mass flow
– % carrier mass flow
– totalizer (mass)