5.4
Error messages
5.4.1
Type of error
Errors which occur during commissioning or measuring operation are displayed immediately. If two
or more system or process errors occur, the error with the highest priority is the one shown on the
display.
The measuring system distinguishes between two types of error:
•
System errors:
This group comprises all device errors, e.g. communication errors, hardware
•
Process errors:
This group comprises all application errors, e.g. inhomogeneous fluid,etc.
a0001211
Fig. 45:
Error messages on the display (example)
1
Error type: P = process error, S = system error
2
Error message type:
$
= fault message, ! = notice message
3
Error designation
4
Error number
5
Duration of most recent error occurrence (hours : minutes : seconds)
5.4.2
Error message type
Error message type
The measuring device always assigns system and process errors which occur to two types of error
messages (
fault
or
notice messages
), resulting in different weightings
Serious system errors, e.g. module defects, are always identified and classed as "fault messages" by
the measuring device.
Notice message (!)
• The error in question has no effect on the current operation.
• Displayed as
→
exclamation mark (!), error type (S: system error, P: process error)
• Displaying the device status on PROFIBUS
Fault message (
$
)
• The error in question interrupts or stops the current operation.
• Displayed as
→
lightning flash (
$
), error type (S: system error, P: process error)
• Displaying the device status on PROFIBUS
1
2
4
5
3
+24.502
X X X X X X X X X X
# 0 0 0 0 0 : 0 0 : 0 5
P
+24.502