Connection with cable extension

To lengthen the sensor connection beyond the length of the fixed cable, you require a junction box VS.

Always connect the sensor cable with the SXP plug to the junction box.
The cable extension to the transmitter then depends from the transmitter version, i.e. field device or panel 
mounted device.

Performance characteristics

Response time

• T

90

: 0.5 minutes

• T

99

: 1.5 minutes (each at 20 °C (68 °F))

Reference operating 
conditions

Signal current in air

1)

approx. 9000 nA

Zero current

zero current free

Measured value resolution

0.001 mg/l (0.001 ppm)

Maximum measured error

±1 % of measured value

2)

Long-term drift

with permanent polarization: < 1 % per month

Influence of medium pressure

Pressure compensation not necessary

Polarization time

< 60 minutes

Oxygen intrinsic consumption

Approx. 900 ng/h in air at 25 °C (77 °F)

a0004129-en

Junction box VS to a field device 

1

SXP plug to field device

2

SXP plug from sensor

a0004130

Junction box VS to a panel mounted device

2

SXP plug from sensor

3

Measuring cable (OMK) to the transmitter

4

Connection department of the transmitter

1

2

82 / 3.23

80

/3.15

mm / inch

3

2

YE
GY
PK
BU
BN
GN
WH

87
0
96
97
88
19
18

4

Reference temperature:

25 °C (77 °F)

Reference pressure:

1013 hPa (15 psi)

1) 

For the reference operating conditions indicated

2) 

In accordance with IEC 746-1 at nominal operating conditions

Summary of Contents for Oxymax W COS71

Page 1: ...itions Laboratory measurement vs continuous process monitoring Spot laboratory measurements to measure oxygen involve a risk of error due to sampling errors Continuous measurement directly in the proc...

Page 2: ...er bromide or silver chloride coating on the anode uses up the bromide or chloride ions dissolved in the electrolyte In the case of conventional membrane covered sensors working with the two electrode...

Page 3: ...el mounting device Remove the SXP connector transmitter side from the cable Refer to the following table for the cable assignment and the assigned terminals for Liquisys M COM223 WX WS Please note tha...

Page 4: ...g l 0 001 ppm Maximum measured error 1 of measured value2 Long term drift with permanent polarization 1 per month Influence of medium pressure Pressure compensation not necessary Polarization time 60...

Page 5: ...ensors You will find them in the Operating Instructions for the assembly used Environment Ambient temperature range 5 to 50 C 20 to 120 F Storage temperature filled with electrolyte 5 to 50 C 20 to 12...

Page 6: ...plug in connection 0 3 kg 0 66 lbs Material Process connection G1 Maximum cable length max 100 m 328 ft including cable extension Membrane thickness approx 25 m Electrolyte Alkaline electrolyte a0004...

Page 7: ...Flow assembly COA260 for plate or wall mounting for oxygen trace measurements ordering acc to product structure Technical Information TI310C 07 en Zero solution 3 units to produce 3 x 1 liter oxygen...

Page 8: ...le Ordering acc to product structure see Technical Information TI 199C 07 en Instruments International Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 71...

Reviews: