Oxymax W COS71
4
Hauser
Connection with cable extension
To lengthen the sensor connection beyond the length of the fixed cable, you require a junction box VS.
Always connect the sensor cable with the SXP plug to the junction box.
The cable extension to the transmitter then depends from the transmitter version, i.e. field device or panel
mounted device.
Performance characteristics
Response time
• T
90
: 0.5 minutes
• T
99
: 1.5 minutes (each at 20 °C (68 °F))
Reference operating
conditions
Signal current in air
1)
approx. 9000 nA
Zero current
zero current free
Measured value resolution
0.001 mg/l (0.001 ppm)
Maximum measured error
±1 % of measured value
2)
Long-term drift
with permanent polarization: < 1 % per month
Influence of medium pressure
Pressure compensation not necessary
Polarization time
< 60 minutes
Oxygen intrinsic consumption
Approx. 900 ng/h in air at 25 °C (77 °F)
a0004129-en
Junction box VS to a field device
1
SXP plug to field device
2
SXP plug from sensor
a0004130
Junction box VS to a panel mounted device
2
SXP plug from sensor
3
Measuring cable (OMK) to the transmitter
4
Connection department of the transmitter
1
2
82 / 3.23
80
/3.15
mm / inch
3
2
YE
GY
PK
BU
BN
GN
WH
87
0
96
97
88
19
18
4
Reference temperature:
25 °C (77 °F)
Reference pressure:
1013 hPa (15 psi)
1)
For the reference operating conditions indicated
2)
In accordance with IEC 746-1 at nominal operating conditions