Oxymax W COS71

Hauser

5

Installation

Angle of installation

The sensor can be installed up to the horizontal in an assembly, support or a suitable process connection.
Other angles are not permissible. Do

 not 

install the sensor overhead.

a0004087

Angle of installation

A

Permissible installation positions: 0 ... 180 °

!

Note! 
Make sure you comply with the instructions for installing sensors. You will find them in the Operating 
Instructions for the assembly used.

Environment

Ambient temperature range

–5 to 50 °C (20 to 120 °F)

Storage temperature

filled with electrolyte: –5 to 50 °C (20 to 120 °F)
without electrolyte: –20 to 60 °C (0 to 140 °F)

Ingress protection

• Fixed cable versions:

IP 68 (10 m (33 ft) water column at 25 °C (77 °F) in 30 days)

• Top 68 plug-in head versions:

IP 68 (1 m (3.3 ft) water column at 50 °C (122 °F) in 7 days)

Process

Process temperature

–5 to 50 °C (20 to 120 °F)

Process pressure

max. 10 bar (145 psi) permissible overpressure
Underpressure operation is not permissible.

A

Summary of Contents for Oxymax W COS71

Page 1: ...itions Laboratory measurement vs continuous process monitoring Spot laboratory measurements to measure oxygen involve a risk of error due to sampling errors Continuous measurement directly in the proc...

Page 2: ...er bromide or silver chloride coating on the anode uses up the bromide or chloride ions dissolved in the electrolyte In the case of conventional membrane covered sensors working with the two electrode...

Page 3: ...el mounting device Remove the SXP connector transmitter side from the cable Refer to the following table for the cable assignment and the assigned terminals for Liquisys M COM223 WX WS Please note tha...

Page 4: ...g l 0 001 ppm Maximum measured error 1 of measured value2 Long term drift with permanent polarization 1 per month Influence of medium pressure Pressure compensation not necessary Polarization time 60...

Page 5: ...ensors You will find them in the Operating Instructions for the assembly used Environment Ambient temperature range 5 to 50 C 20 to 120 F Storage temperature filled with electrolyte 5 to 50 C 20 to 12...

Page 6: ...plug in connection 0 3 kg 0 66 lbs Material Process connection G1 Maximum cable length max 100 m 328 ft including cable extension Membrane thickness approx 25 m Electrolyte Alkaline electrolyte a0004...

Page 7: ...Flow assembly COA260 for plate or wall mounting for oxygen trace measurements ordering acc to product structure Technical Information TI310C 07 en Zero solution 3 units to produce 3 x 1 liter oxygen...

Page 8: ...le Ordering acc to product structure see Technical Information TI 199C 07 en Instruments International Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 71...

Reviews: