background image

Oxymax W COS71

2

Hauser

Function and system design

Measuring principle

The oxygen molecules diffused through the membrane are reduced to hydroxide ions (OH-) at the cathode. 
Silver is oxidized to silver ions (Ag+) at the anode (this forms a silver halogenide layer).
A current flows due to the electron donation at the cathode and the electron acceptance at the anode. Under 
constant conditions, this flow is proportional to the oxygen content of the medium.
This current is converted in the transmitter and indicated on the display as an oxygen concentration in mg/l, 
as a saturation index in % SAT or as an oxygen partial pressure in hPa.

Potentiostatic amperometric 
three-electrode system

The high-impedance, current-free reference electrode plays an important role.
The formation of a silver bromide or silver chloride coating on the anode uses up the bromide or chloride ions 
dissolved in the electrolyte
In the case of conventional membrane-covered sensors working with the two-electrode system, this causes an 
increase in signal drift.

This is not the case with the three-electrode system:
The change in bromide or chloride concentration is registered by the reference electrode and an internal control 
circuit holds the working electrode potential constant. The advantages of this principle are significantly 
increased accuracy of the signal and considerably extended calibration intervals.

Measuring system

A complete measuring system comprises at least:
• Oxygen sensor
• Transmitter, e.g. Liquisys M COM223/253-WX/WS
• Special measuring cable
• Assembly, e.g. flow assembly COA260 or retractable assembly COA451

Optional:
• Junction box VS (with cable extension)

a0004102

Measuring system (example)

1

Flow assembly

2

Junction box VS (optional)

3

Transmitter Liquisys M COM253

4

Oxygen sensor

Summary of Contents for Oxymax W COS71

Page 1: ...itions Laboratory measurement vs continuous process monitoring Spot laboratory measurements to measure oxygen involve a risk of error due to sampling errors Continuous measurement directly in the proc...

Page 2: ...er bromide or silver chloride coating on the anode uses up the bromide or chloride ions dissolved in the electrolyte In the case of conventional membrane covered sensors working with the two electrode...

Page 3: ...el mounting device Remove the SXP connector transmitter side from the cable Refer to the following table for the cable assignment and the assigned terminals for Liquisys M COM223 WX WS Please note tha...

Page 4: ...g l 0 001 ppm Maximum measured error 1 of measured value2 Long term drift with permanent polarization 1 per month Influence of medium pressure Pressure compensation not necessary Polarization time 60...

Page 5: ...ensors You will find them in the Operating Instructions for the assembly used Environment Ambient temperature range 5 to 50 C 20 to 120 F Storage temperature filled with electrolyte 5 to 50 C 20 to 12...

Page 6: ...plug in connection 0 3 kg 0 66 lbs Material Process connection G1 Maximum cable length max 100 m 328 ft including cable extension Membrane thickness approx 25 m Electrolyte Alkaline electrolyte a0004...

Page 7: ...Flow assembly COA260 for plate or wall mounting for oxygen trace measurements ordering acc to product structure Technical Information TI310C 07 en Zero solution 3 units to produce 3 x 1 liter oxygen...

Page 8: ...le Ordering acc to product structure see Technical Information TI 199C 07 en Instruments International Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 71...

Reviews: