Commissioning

Hauser

7

3.2

Post-connection check

4

Commissioning

4.1

Function check

Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.

If using an assembly with automatic cleaning, check the correct water connection at the 
assembly rinse connection.

#

Warning! 
Danger of medium leaking off
Before applying compressed air to an assembly with cleaning facility, make sure the connections 
are correctly fitted. Otherwise, the assembly may not be insert into the process.

4.2

Polarization

The sensor was tested in the factory for perfect functionality and is supplied ready for operation.

To prepare for calibration, proceed as follows:

1.

Remove the sensor protective cap.

2.

Place the externally dry sensor in atmospheric air. The air should be saturated with water 
vapour. Therefore, install the sensor as close to the water surface as possible. When 
calibrating the sensor membrane, make sure the membrane remains dry. Therefore, avoid 
any direct contact with the water surface.

3.

Connect the sensor to the transmitter and switch on the transmitter.

Instrument status and specifications

Remarks

Are the sensor, assembly, junction box or cable damaged?

Visual inspection

Electrical connection

Remarks

Are the installed cables strain-relieved and not twisted ?

Is the cable type route completely isolated ?

Power cable/weak current cable

Are the power supply and signal cable correctly connected to the 
transmitter ?

Use the connection diagram of the transmitter.

Are all the screws terminals properly tightened ?

Tighten

Are all the cable entries installed, tightened and sealed ?

For cable entries lateral: cable loops downwards 
for water to be able to drip off.

Are all the cable entries installed downwards or lateral ?

Summary of Contents for Oxymax H COS21

Page 1: ...BA244C 07 en 09 06 71032330 Operating Instructions Oxymax H COS21 Dissolved oxygen sensor...

Page 2: ...tified electrician Technical personnel must have read and understood these Operating Instructions and must adhere to them Before commissioning the entire measuring point check all the connections for...

Page 3: ...e contents are undamaged Inform the supplier about damage to the delivery contents Keep the damaged products until the matter has been settled Check that the scope of delivery is complete and agrees w...

Page 4: ...2 2 Mounting location Select the installation location so that there is easy access for later calibration Make sure that upright posts and assemblies are secured safely and vibration free For immerse...

Page 5: ...ll and connect the oxygen sensor Caution Do not install the sensor suspended from the cable Screw the sensor into the assembly so that the cable is not twisted Avoid exerting excessive tensile force o...

Page 6: ...Wiring Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them...

Page 7: ...eric air The air should be saturated with water vapour Therefore install the sensor as close to the water surface as possible When calibrating the sensor membrane make sure the membrane remains dry Th...

Page 8: ...peration by taking it out of the fluid drying it and then measuring the oxygen saturation index at air after 10 minutes Decide using the results a If the measured value is not at 102 2 SAT you have to...

Page 9: ...ing proceed as follows Caution After cleaning rinse the sensor with copious amounts of clean water 5 2 Regenerating the sensor 5 2 1 Replacing the sealing ring Replacing the sealing ring is only neces...

Page 10: ...the membrane cap 4 If necessary clean the cathode or replace the sealing ring if it is damaged 5 Rinse the electrode holder with drinking water Installing the new membrane cap 1 Make sure that there a...

Page 11: ......

Page 12: ...www endress com worldwide BA244C 07 en 09 06 Printed in Germany FM SGML 6 0 DT 71032330...

Reviews: