Wiring

Oxymax H COS21

6

Hauser

#

Warning! 
When using metallic assemblies and installation equipment, comply with national grounding 
regulations.

2.4

Post-installation check

• Sensor and cable undamaged?
• Cap undamaged?
• Compliance with permissible sensor installation position?
• Is the sensor installed in an assembly and is not suspended from the cable?
• Avoid moisture by rain by fitting the protective cap to the immersion assembly?

3

Wiring

#

Warning! 
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must 

adhere to them.

• Ensure that there is no voltage at the power cable before beginning the connection work.

3.1

Connection to Liquisys M COM223/253 F

The electrical connection between the sensor and the transmitter uses the multi-core special 
measuring cable COK21.

a0005583-en

Fig. 3:  Special measuring cable COK21

PK

GN / YE

BN

YE

BU

S

12

90

91

11

Assignment

Terminal

External shield

Cathode

Anode

NTC temp. sensor

Active internal shield

(NTC temp. sensor)

Summary of Contents for Oxymax H COS21

Page 1: ...BA244C 07 en 09 06 71032330 Operating Instructions Oxymax H COS21 Dissolved oxygen sensor...

Page 2: ...tified electrician Technical personnel must have read and understood these Operating Instructions and must adhere to them Before commissioning the entire measuring point check all the connections for...

Page 3: ...e contents are undamaged Inform the supplier about damage to the delivery contents Keep the damaged products until the matter has been settled Check that the scope of delivery is complete and agrees w...

Page 4: ...2 2 Mounting location Select the installation location so that there is easy access for later calibration Make sure that upright posts and assemblies are secured safely and vibration free For immerse...

Page 5: ...ll and connect the oxygen sensor Caution Do not install the sensor suspended from the cable Screw the sensor into the assembly so that the cable is not twisted Avoid exerting excessive tensile force o...

Page 6: ...Wiring Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them...

Page 7: ...eric air The air should be saturated with water vapour Therefore install the sensor as close to the water surface as possible When calibrating the sensor membrane make sure the membrane remains dry Th...

Page 8: ...peration by taking it out of the fluid drying it and then measuring the oxygen saturation index at air after 10 minutes Decide using the results a If the measured value is not at 102 2 SAT you have to...

Page 9: ...ing proceed as follows Caution After cleaning rinse the sensor with copious amounts of clean water 5 2 Regenerating the sensor 5 2 1 Replacing the sealing ring Replacing the sealing ring is only neces...

Page 10: ...the membrane cap 4 If necessary clean the cathode or replace the sealing ring if it is damaged 5 Rinse the electrode holder with drinking water Installing the new membrane cap 1 Make sure that there a...

Page 11: ......

Page 12: ...www endress com worldwide BA244C 07 en 09 06 Printed in Germany FM SGML 6 0 DT 71032330...

Reviews: