Diagnostics and troubleshooting
LPGmass Modbus RS485
64
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
832
Electronic temperature too
high
Reduce ambient temperature
S
Warning
833
Electronic temperature too
low
Increase ambient temperature
S
Warning
834
Process temperature too
high
Reduce process temperature
S
Warning
835
Process temperature too
low
Increase process temperature
S
Warning
843
Process limit
Check process conditions
S
Warning
862
Partly filled pipe
1. Check for gas in process
2. Adjust detection limits
S
Warning
910
Tubes not oscillating
1. Check electronic
2. Inspect sensor
F
Alarm
912
Medium inhomogeneous
1. Check process cond.
2. Increase system pressure
S
Warning
912
Inhomogeneous
S
Warning
913
Medium unsuitable
1. Check process conditions
2. Check electronic modules or
sensor
S
Alarm
944
Monitoring failed
Check process conditions for
Heartbeat Monitoring
S
Warning
948
Tube damping too high
Check process conditions
S
Warning
990
Special event 4
Contact service
F
Alarm
991
Special event 8
Contact service
F
Alarm
992
Special event 12
Contact service
F
Alarm
1)
Diagnostic behavior can be changed.
12.7 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 61
• Via "DeviceCare" operating tool → 61
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
Previous diagnostics