Micropilot FMR62
42
V. 4, Rev. 2, 19-03-2018
Hauser
FMR62, flange or DIN11851 process connection, PTFE flush mount
p
-40
(-40)
+150
(+302)
+200
(+392)
0
(+32)
T
p
[°C]
([°F])
1
2
[bar] ([psi])
25 (362.6)
-1 (-14.5)
0 (0)
A0032184
14
FMR62: Permitted range for process temperature and process pressure, flange or DIN11851 process
connection
1
Feature 90, seal: F5
2
Feature 90, seal: F6
FMR62, flange or DIN11851 process connection
Feature 100 "Process connection"
Feature
90 "Seal"
Process
temperature
range
Process pressure range
• AFK:
NPS 2" Cl.150 RF, PTFE > 316/316L
• AGK:
NPS 3" Cl.150 RF, PTFE > 316/316L
• AHK:
NPS 4" Cl.150 RF, PTFE > 316/316L
• AJK:
NPS 6" Cl.150 RF, PTFE > 316/316L
• CFK:
DN50 PN10/16 B1, PTFE > 316L
• CGK:
DN80 PN10/16 B1, PTFE > 316L
• CHK:
DN100 PN10/16 B1, PTFE > 316L
• CJK:
DN150 PN10/16 B1, PTFE > 316L
• KFK:
10K 50A RF, PTFE > 316L
• KGK:
10K 80A RF, PTFE > 316L
• KHK:
10K 100A RF, PTFE > 316L
• KJK:
10K 150A RF, PTFE > 316L
• MRK:
DIN11851 DN50 PN25 slotted nut, PTFE > 316L
• MTK:
DIN11851 DN80 PN25 slotted nut, PTFE > 316L
F5:
PTFE
cladded
–40 to +150 °C
(–40 to +302 °F)
p
rel
=
–1 to 25 bar
(–14.5 to 362.6 psi)
F6:
PTFE
cladded
–40 to +200 °C
(–40 to +392 °F)
1)
The pressure range may be further restricted in the event of a CRN approval
Dielectric constant
For liquids
ε
r
≥ 1.9
Please contact Hauser for applications with lower dielectric constants than indicated.
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Hauser DC manual (CP01076F)
• the Hauser "DC Values App" (available for Android and iOS)