Micropilot FMR62
34
V. 4, Rev. 2, 19-03-2018
Hauser
Inner nozzle diameter
D
Maximum nozzle height
H
max
Antenna GM
:
50mm / 2"
Antenna GN
80mm / 3"
100 to 150 mm (4 to 6 in)
1
250 mm (50 in)
2
200 mm (88 in)
≥ 150 mm (6 in)
1
850 mm (74 in)
3
300 mm (132 in)
1)
In case of longer nozzles, a reduction of the measuring performance is to be expected.
2)
Merkamal 070 der Produktstruktur
Note the following if the antenna does not project out of the nozzle:
• The end of the nozzle must be smooth and free from burrs. The edge of the nozzle should be
rounded if possible.
• Mapping must be performed.
• Please contact Hauser for applications with nozzles that are higher than indicated in
the table.
Mounting cladded flanges
Note the following for cladded flanges:
• Use the same number of flange screws as the number of flange bores provided.
• Tighten the screws with the necessary torque (see Table).
• Retighten after 24 hours or after the first temperature cycle.
• Depending on the process pressure and temperature, check and retighten the screws, where
necessary, at regular intervals.
The PTFE flange cladding normally acts simultaneously as a seal between the nozzle and the
device flange.
Flange size
Number of screws
Recommended tightening torque [Nm]
Minimum
Maximum
EN
DN50/PN16
4
45
65
DN80/PN16
8
40
55
DN100/PN16
8
40
60
DN150/PN16
8
75
115
ASME
2"/150lbs
4
40
55
3"/150lbs
4
65
95
4"/150lbs
8
45
70
6"/150lbs
8
85
125
JIS
10K 50A
4
40
60
10K 80A
8
25
35
10K 100A
8
35
55
10K 150A
8
75
115