background image

Summary of Contents for Micropilot FMR62

Page 1: ...n in the case of small vessels and small process connections Safety by design ensures highest safety Easy guided commissioning with intuitive user interface Bluetooth wireless technology for commissioning operation and maintenance via free iOS Android app SmartBlue Maximum reliability thanks to multi echo tracking HistoROM configuration memory makes for easier commissioning maintenance and diagnos...

Page 2: ...erature limits 36 Storage temperature 37 Climate class 37 Altitude according to IEC61010 1 Ed 3 37 Degree of protection 37 Vibration resistance 38 Electromagnetic compatibility EMC 38 Process 39 Process temperature process pressure 39 Dielectric constant 42 Mechanical construction 43 Dimensions 43 Weight 50 Materials GT18 housing stainless steel corrosion resistant 51 Materials GT19 housing plasti...

Page 3: ...Application Packages 75 Heartbeat Diagnostics 75 Heartbeat Verification 76 Heartbeat Monitoring 77 Accessories 78 Device specific accessories 78 Communication specific accessories 81 Service specific accessories 82 System components 82 Supplementary documentation 83 Standard documentation 83 Safety Instructions XA 83 ...

Page 4: ...rrent and alternating current Ground connection A grounded terminal which as far as the operator is concerned is grounded via a grounding system Protective Earth PE A terminal which must be connected to ground prior to establishing any other connections The ground terminals are situated inside and outside the device Inner ground terminal Connects the protectiv earth to the mains supply Outer groun...

Page 5: ... Safe area non hazardous area Indicates the non hazardous area Symbols at the device Symbol Meaning Safety instructions Observe the safety instructions contained in the associated Operating Instructions Temperature resistance of the connection cables Specifies the minimum value of the temperature resistance of the connection cables ...

Page 6: ...he nameplate ToF Time of Flight FieldCare Scalable software tool for device configuration and integrated plant asset management solutions DeviceCare Universal configuration software for Endress Hauser HART PROFIBUS FOUNDATION Fieldbus and Ethernet field devices DTM Device Type Manager DD Device Description for HART communication protocol DC Relative dielectric constant εr Operating tool The term o...

Page 7: ...ademarks and trade names are those of their respective owners Apple Apple the Apple logo iPhone and iPod touch are trademarks of Apple Inc registered in the U S and other countries App Store is a service mark of Apple Inc Android Android Google Play and the Google Play logo are trademarks of Google Inc KALREZ VITON Registered trademark of DuPont Performance Elastomers L L C Wilmington USA TEFLON R...

Page 8: ...and product surface The frequency of this wave is modulated in the form of a sawtooth signal between two limit frequencies f1 and f2 f t 1 2 f2 f1 t f A0023771 2 FMCW principle result of frequency modulation 1 Transmitted signal 2 Received signal This results in the following difference frequency at any time between the transmitted signal and the received signal Δf k Δt where Δt is the run time an...

Page 9: ...ective properties the installation position and any possible interference reflections The following tables describe the groups of media as well as the achievable measuring range as a function of application and media group If the dielectric constant of a medium is unknown it is recommended to assume media group B to ensure a reliable measurement Media groups Media group DC εr Example A0 1 2 to 1 4...

Page 10: ...rmittent filling filling from bottom immersion tubes A0034566 Moving surfaces e g continuous filling from above mixing jets A0034567 Turbulent surface e g filling from above agitators baffles A0 A B C D 1 5 5 2 5 8 5 16 8 26 10 33 A0034575 A0 A B C D 1 3 3 1 5 5 3 10 6 20 8 26 A0034580 A B C D 1 3 3 1 5 5 3 10 5 16 A0034585 Measuring range m ft 1 Product structure feature 070 option GE ...

Page 11: ...tent filling filling from bottom immersion tubes A0034566 Moving surfaces e g continuous filling from above mixing jets A0034567 Turbulent surface e g filling from above agitators baffles A0 A B C D 3 10 6 20 11 36 15 49 22 72 A0034576 A0 A B C D 1 5 5 3 10 6 20 13 43 20 66 A0034581 A0 A B C D 1 3 3 1 5 5 3 10 7 23 11 36 A0034586 Measuring range m ft 1 Product structure feature 070 option GF ...

Page 12: ...ittent filling filling from bottom immersion tubes A0034566 Moving surfaces e g continuous filling from above mixing jets A0034567 Turbulent surface e g filling from above agitators baffles A0 A B C D 7 23 12 39 23 75 40 131 50 164 A0034577 A0 A B C D 4 13 7 23 13 43 28 92 44 144 A0034582 A0 A B C D 2 7 4 13 7 23 15 49 25 82 A0034587 Measuring range m ft 1 Product structure feature 070 option GM ...

Page 13: ... surfaces e g continuous filling from above mixing jets A0034567 Turbulent surface e g filling from above agitators baffles A0 A B C D 22 72 40 131 50 164 65 213 80 262 A0034578 A0 A B C D 12 39 23 75 45 148 60 197 70 230 A0034583 A0 A B C D 7 23 13 43 25 82 50 164 60 197 A0034588 Measuring range m ft 1 Product structure feature 070 option GN Operating frequency Approx 80 GHz Up to 8 devices can b...

Page 14: ...Micropilot FMR62 14 V 4 Rev 2 19 03 2018 Endress Hauser Transmission power Peak power 6 3 mW Average output power 63 µW ...

Page 15: ...Binary conductive or non conductive switches when the programmable switch point is reached Failure mode non conductive Electrical connection values U 16 to 35 VDC I 0 to 40 mA Internal resistance RI 880 Ω The voltage drop at this internal resistance has to be taken into account on planning the configuration For example the resulting voltage at a connected relay must be sufficient to switch the rel...

Page 16: ...e entered manually or semi automatically Galvanic isolation All circuits for the outputs are galvanically isolated from each other Protocol specific data HART Manufacturer ID 17 0x11 Device type ID 0x112B HART specification 7 0 Device description files DTM DD Information and files under www endress com www fieldcommgroup org HART load min 250 Ω HART device variables The measured values can be free...

Page 17: ...Micropilot FMR62 Endress Hauser V 4 Rev 2 19 03 2018 17 Multidrop current 4 0 mA Set up time 15 s ...

Page 18: ...re 4 20mA HART A Without integrated overvoltage protection B With integrated overvoltage protection 1 Active barrier with power supply e g RN221N Observe terminal voltage 2 HART communication resistor 250 Ω Observe maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350 SFX370 via VIATOR Bluetooth modem 4 Analog display device Observe maximum load 5 Cable screen observe cable specificat...

Page 19: ...d overvoltage protection B With integrated overvoltage protection 1 Active barrier with power supply e g RN221N Observe terminal voltage 2 HART communication resistor 250 Ω Observe maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350 SFX370 via VIATOR Bluetooth modem 4 Analog display device Observe maximum load 5 Cable screen observe cable specification 6 4 20mA HART passive Terminal...

Page 20: ...output 1 3 Supply voltage for current output 1 e g RN221N Observe terminal voltage 4 Cable screen observe cable specification 5 HART communication resistor 250 Ω Observe maximum load 6 Connection for Commubox FXA195 or FieldXpert SFX350 SFX370 via VIATOR Bluetooth modem 7 Analog display device observe maximum load 8 Analog display device observe maximum load 9 Supply voltage for current output 2 e...

Page 21: ...utput 3 4 Ri A0015909 6 Connection of a relay Suitable relays examples Solid state relay Phoenix Contact OV 24DC 480AC 5 with mounting rail connector UMK 1 OM R AMS Electromechanical relay Phoenix Contact PLC RSC 12DC 21 3 2 1 4 Ri A0015910 7 Connection of a digital input 1 Pull up resistor 2 Digital input For optimum interference immunity we recommend to connect an external resistor internal resi...

Page 22: ...plug connector M12 or 7 8 the signal line can be connected without opening the housing Pin assignment of the M12 plug connector 2 1 3 4 A0011175 Pin Meaning 1 Signal 2 not connected 3 Signal 4 Ground Pin assignment of the 7 8 plug connector 4 2 3 1 A0011176 Pin Meaning 1 Signal 2 Signal 3 Not connected 4 Screen ...

Page 23: ...eature 020 in the product structure 2 Feature 010 in the product structure 3 If the Bluetooth modem is used the minimum supply voltage increases by 2 V 4 At ambient temperatures TTa 20 C a terminal voltage U 16 V is required to start the device with the min error current 3 6 mA Power supply output 1 Approval 2 Terminal voltage U at device Maximum load R depending on the supply voltage U0 of the po...

Page 24: ...th f 100 to 10000 Hz USS 10 mV Power consumption Power supply Output 1 Power consumption A 2 wire 4 20mA HART 0 9 W B 2 wire 4 20mA HART switch output 0 9 W C 2 wire 4 20mA HART 4 20mA 2 x 0 7 W 1 Feature 020 of the product structure Current consumption HART Nominal current 3 6 to 22 mA the start up current for multidrop mode can be parametrized is set to 3 6 mA on delivery Breakdown signal NAMUR ...

Page 25: ...60 C 140 F use cable for temperature TU 20 K HART A normal device cable suffices if only the analog signal is used A shielded cable is recommended if using the HART protocol Observe grounding concept of the plant Overvoltage protection If the measuring device is used for level measurement in flammable liquids which requires the use of overvoltage protection according to DIN EN 60079 14 standard fo...

Page 26: ...nting conditions may be up to 4 mm 0 16 in This additional offset zero point can be compensated for by entering a correction parameter level correction during commissioning 3 The non repeatability is already considered in the accuracy Differing values in near range applications 0 8 2 62 4 0 16 1 0 04 1 0 04 4 0 16 D m mm in R 0 A0032636 8 Maximum measured error in near range applications Δ Maximum...

Page 27: ...s in a systematic measured error that increases with increasing distance between the reference point of the measurement flange and the surface of the product The following table shows this measured error for a few typical gases vapors with regard to the distance a positive value means that a too large distance is measured Gas phase Temperature Pressure C F 1 bar 14 5 psi 10 bar 145 psi 25 bar 362 ...

Page 28: ...l nozzle outer edge 1 6 of the container diameter However the device must not under any circumstances be mounted closer than 15 cm 5 91 in to the tank wall Not in the center 2 as interference can cause signal loss Not above the filling curtain 3 The use of a weather protection cover 1 is recommended to protect the transmitter from direct sunlight or rain ...

Page 29: ...Rev 2 19 03 2018 29 Internal container fittings α A0031777 Avoid the location of internal fittings limit switches temperature sensors struts vacuum rings heating coils baffles etc inside the signal beam Take into account the beam angle 30 ...

Page 30: ...esulting in fewer interference echoes 30 Mapping Measurement can be optimized by electronically suppressing interference echoes Beam angle D W D _ 2 2 tan W α α A0031824 9 Relationship between beam angle α distance D and beamwidth diameter W The beam angle is defined as the angle α where the energy density of the radar waves reaches half the value of the maximum energy density 3dB width Microwaves...

Page 31: ...63 m 8 63 ft 1 40 m 4 58 ft 1 22 m 4 00 ft 0 52 m 1 71 ft 15 m 49 ft 2 09 m 6 87 ft 1 83 m 6 01 ft 0 79 m 2 59 ft 20 m 66 ft 2 79 m 9 16 ft 2 44 m 8 01 ft 1 05 m 3 44 ft 25 m 82 ft 3 05 m 10 02 ft 1 31 m 4 30 ft 30 m 98 ft 3 66 m 12 02 ft 1 57 m 5 15 ft 35 m 115 ft 4 27 m 14 02 ft 1 83 m 6 00 ft 40 m 131 ft 4 88 m 16 03 ft 2 09 m 6 86 ft 45 m 148 ft 5 50 m 18 03 ft 2 36 m 7 74 ft 50 m 164 ft 6 11 ...

Page 32: ... might occur In case of outdoor mounting the space between antenna and vessel has to be protected from the elements Do not mount any potential reflectors e g pipes outside the tank in the signal beam Suitable thickness of the tank ceiling or dielectric window Material PE PTFE PP Perspex DK εr 2 3 2 1 2 3 3 1 Optimum thickness 1 25 mm 0 049 in 1 1 3 mm 0 051 1 1 25 mm 0 049 in 1 1 07 mm 0 042 in 1 ...

Page 33: ...st be smooth and free from burrs The edge of the nozzle should be rounded if possible Mapping must be performed Please contact Endress Hauser for applications with nozzles that are higher than indicated in the table Information concerning threaded connections When screwing in turn by the hex bolt only Tool open ended wrench 36 mm Sensor 3 4 open ended wrench 55 mm Sensor 1 1 2 Maximum permissible ...

Page 34: ...es that are higher than indicated in the table Mounting cladded flanges Note the following for cladded flanges Use the same number of flange screws as the number of flange bores provided Tighten the screws with the necessary torque see Table Retighten after 24 hours or after the first temperature cycle Depending on the process pressure and temperature check and retighten the screws where necessary...

Page 35: ...sulation 1 2 A0032207 If process temperatures are high the device should be included in the usual container insulation system 2 to prevent the electronics from heating as a result of thermal radiation or convection The insulation should not be higher than the neck of the device 1 ...

Page 36: ...ambient temperature Ta is reduced as indicated in the following diagram temperature derating in the table header FMR62 Antenna 1 GE Integrated PEEK 3 4 GF Integrated PEEK 1 1 2 GM PTFE cladded flush mount DN50 GN PTFE cladded flush mount DN80 P2 P1 P5 P4 P3 Ta Tp A0032024 Seal 2 A6 FKM Viton GLT 40 to 200 C 40 to 392 F C2 FFKM Kalrez 20 to 200 C 4 to 392 F F6 PTFE cladded 40 to 200 C 40 to 392 F T...

Page 37: ...oduct structure 2 Feature 090 in the product structure 3 Feature 040 in the product structure Storage temperature 40 to 80 C 40 to 176 F Climate class DIN EN 60068 2 38 test Z AD Altitude according to IEC61010 1 Ed 3 Generally up to 2000 m 6600 ft above MSL Above 2000 m 6600 ft if the following conditions are met Ordering feature 020 Power supply Output A B C E or G 2 wire versions Supply voltage ...

Page 38: ...200 C 40 to 392 F incl gas tight feedthrough Electromagnetic compatibility EMC Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the EN 61326 series and NAMUR Recommendation EMC NE 21 For details please refer to the Declaration of Conformity 4 A normal device cable suffices if only the analog signal should be used Use a shielded cable for digital communi...

Page 39: ...entical ASME B 16 5a 1998 Tab 2 2 2 F316 ASME B 16 5a 1998 Tab 2 3 8 N10276 JIS B 2220 FMR62 integrated antenna PEEK 150 C 302 F p 40 40 20 4 150 302 0 32 Tp C F 1 2 bar psi 20 290 1 14 5 0 0 A0032182 10 FMR62 Permitted range for process temperature and process pressure integrated antenna PEEK feature 070 GE GF 150 C 302 F 1 Feature 90 seal A5 FKM Viton GLT 2 Feature 90 seal C1 FFKM Kalrez FMR62 i...

Page 40: ...ection Feature 90 Seal Process temperature range Process pressure range GDJ Thread ISO228 G3 4 316L GGJ Thread ISO228 G1 1 2 316L RDJ Thread ANSI MNPT3 4 316L RGJ Thread ANSI MNPT1 1 2 316L A6 FKM Viton GLT 40 to 200 C 40 to 392 F prel 1 to 20 bar 14 5 to 290 psi 1 C2 FFKM Kalrez 20 to 200 C 4 to 392 F 1 The pressure range may be further restricted in the event of a CRN approval FMR62 Tri Clamp 2 ...

Page 41: ...ss connection p 40 40 150 302 200 392 0 32 Tp C F 1 2 bar psi 14 203 1 14 5 0 0 A0032619 13 FMR62 Permitted range for process temperature and process pressure Tri Clamp 3 or Tri Clamp 4 process connection 1 Feature 90 seal F5 2 Feature 90 seal F6 FMR62 Tri Clamp 3 or Tri Clamp 4 process connection Feature 100 Process connection Feature 90 Seal Process temperature range Process pressure range TFK T...

Page 42: ...0 PN10 16 B1 PTFE 316L CGK DN80 PN10 16 B1 PTFE 316L CHK DN100 PN10 16 B1 PTFE 316L CJK DN150 PN10 16 B1 PTFE 316L KFK 10K 50A RF PTFE 316L KGK 10K 80A RF PTFE 316L KHK 10K 100A RF PTFE 316L KJK 10K 150A RF PTFE 316L MRK DIN11851 DN50 PN25 slotted nut PTFE 316L MTK DIN11851 DN80 PN25 slotted nut PTFE 316L F5 PTFE cladded 40 to 150 C 40 to 302 F prel 1 to 25 bar 14 5 to 362 6 psi 1 F6 PTFE cladded ...

Page 43: ...ensions in mm in for devices with integrated overvoltage protection 144 5 67 113 5 4 47 ø106 4 17 78 3 07 90 3 54 99 3 9 ø 1 0 6 4 1 7 R110 A0011346 16 Housing GT19 Plastics PBT Dimensions in mm in for devices with integrated overvoltage protection 145 5 5 73 141 5 5 57 ø112 4 41 ø108 5 4 27 78 3 07 90 3 54 98 3 86 ø 1 0 3 5 4 0 7 R110 61 2 4 A0020751 17 Housing GT20 Alu coated Dimensions in mm in...

Page 44: ...ed through C1 FFKM Kalrez 20 150 C 4 302 F C2 FFKM Kalrez 20 200 C 4 392 F incl gas tight feed through 100 Process Connection GDJ Thread ISO228 G3 4 316L RDJ Thread ANSI MNPT3 4 316L ø23 8 0 94 143 5 63 186 7 32 22 0 87 1 A ø32 2 1 27 B 1 R A0031557 18 Dimensions mm in A Seal FKM Viton GLT 40 to 150 C 40 to 302 F or FFKM Kalrez 20 150oC 4 302 F B Seal FKM Viton GLT 40 to 200 C 40 to 392 F or FFKM ...

Page 45: ...ed through C1 FFKM Kalrez 20 150 C 4 302 F C2 FFKM Kalrez 20 200 C 4 392 F incl gas tight feed through 100 Process Connection GGJ Thread ISO228 G1 1 2 316L RGJ Thread ANSI MNPT1 1 2 316L 227 8 94 A 30 1 18 184 7 24 B ø42 5 1 67 ø54 8 2 18 R 1 1 A0032787 19 Dimensions mm in A Seal FKM Viton GLT 40 to 150 C 40 to 302 F or FFKM Kalrez 20 150oC 4 302 F B Seal FKM Viton GLT 40 to 200 C 40 to 392 F or F...

Page 46: ...K DIN11851 DN50 PN25 slotted nut PTFE 316L MTK DIN11851 DN80 PN25 slotted nut PTFE 316L L1 L3 øD2 øD1 L2 L4 1 A B R 1 A0032175 20 Dimensions mm in A Seal PTFE cladded 40 to 150 C 40 to 302 F B Seal PTFE cladded 40 to 200 C 40 to 392 F R Reference point of the measurement 1 Bottom edge of housing Feature 100 Process connection øD1 øD2 L1 L2 L3 L4 MRK DN50 PN25 slotted nut PTFE 316L 48 mm 1 89 in 65...

Page 47: ... 316L TDK Tri Clamp ISO2852 DN101 6 4 PTFE 316L øD2 øD1 L1 L2 L4 1 1 øD3 2 6 0 10 1 6 0 06 A B R A0032176 21 Dimensions mm in A Seal PTFE cladded 40 to 150 C 40 to 302 F B Seal PTFE cladded 40 to 200 C 40 to 392 F R Reference point of the measurement 1 Bottom edge of housing Feature 70 Antenna øD1 L1 L2 L4 GM PTFE cladded flush mount DN50 48 mm 1 89 in 145 mm 5 71 in 188 mm 7 40 in 19 mm 0 75 in G...

Page 48: ...4 Cl 300 PTFE 316 316L flange ASME B16 5 CFK DN50 PN10 16 PTFE 316L flange EN1092 1 CGK DN80 PN10 16 PTFE 316L flange EN1092 1 CHK DN100 PN10 16 PTFE 316L flange EN1092 1 CJK DN150 PN10 16 PTFE 316L flange EN1092 1 KFK 10K 50A PTFE 316L flange JIS B2220 KGK 10K 80A PTFE 316L flange JIS B2220 KHK 10K 100A PTFE 316L Flansch JIS B2220 KJK 10K 150A PTFE 316L flange JIS B2220 L3 L4 øD2 øD3 øD1 L1 4 0 1...

Page 49: ...4 Cl 300 PTFE 316 316L 158 mm 6 2 in 255 mm 10 in 32 mm 1 3 in CFK DN50 PN10 16 PTFE 316L 102 mm 4 02 in 165 mm 6 50 in 18 mm 0 71 in CGK DN80 PN10 16 PTFE 316L 138 mm 5 43 in 200 mm 7 87 in 20 mm 0 79 in CHK DN100 PN10 16 PTFE 316L 158 mm 6 22 in 220 mm 8 66 in 20 mm 0 79 in CJK DN150 PN10 16 PTFE 316L 212 mm 8 35 in 285 mm 11 22 in 22 mm 0 87 in KFK 10K 50A PTFE 316L 96 mm 3 78 in 155 mm 6 10 in...

Page 50: ...enna and process connection Device Antenna 1 Weight of antenna process connection FMR62 GE Integrated PEEK 3 4 Max 1 5 kg 3 31 lb flange weight 2 GF Integrated PEEK 1 1 2 Max 2 3 kg 5 07 lb flange weight 2 GM PTFE cladded flush mount DN50 Max 1 7 kg 3 75 lb flange weight 2 GN PTFE cladded flush mount DN80 Max 2 9 kg 6 39 lb flange weight 2 1 Order code 070 2 For flange weights 316 316L see Technic...

Page 51: ...lug cable gland adapter or plug depending on the device version Dummy plug depending on the device version PE PBT GF Cable gland 316L 1 4404 or nickel plated brass Adapter 316L 1 4404 1 4435 Seal EPDM M12 plug Nickel plated brass 1 7 8 plug 316 1 4401 2 5 2 Dummy plug cable gland or adapter depending on the device version Dummy plug 316L 1 4404 Cable gland 316L 1 4404 or nickel plated brass Adapte...

Page 52: ...pending on the device version Nickel plated brass CuZn PA Adapter 316L 1 4404 1 4435 Seal EPDM M12 plug Nickel plated brass 1 7 8 plug 316 1 4401 2 5 2 Dummy plug cable gland or adapter depending on the device version Dummy plug depending on the device version PE PBT GF Nickel plated steel Cable gland depending on the device version Nickel plated brass CuZn PA Adapter 316L 1 4404 1 4435 Seal EPDM ...

Page 53: ... 4404 4 Lock at the housing neck Screw A4 70 Clamp 316L 1 4404 5 1 Dummy plug cable gland adapter or plug depending on the device version Dummy plug depending on the device version PE PBT GF Cable gland depending on the device version Nickel plated brass CuZn PA Adapter 316L 1 4404 1 4435 Seal EPDM M12 plug Nickel plated brass 1 7 8 plug 316 1 4401 2 5 2 Dummy plug cable gland or adapter depending...

Page 54: ... Materials antenna and process connection FMR62 integrated antenna 1 2 3 4 5 A0032019 No Component part Material 1 Antenna Antenna PEEK Antenna seal Viton FKM KALREZ FFKM 2 Antenna adapter 316L 1 4404 3 Housing adapter 316L 1 4404 4 Housing seal EPDM 5 Process connection 316L 1 4404 FMR62 antenna cladded flush mount 1 2 3 4 5 6 1 2 3 4 5 1 2 3 4 5 A0031815 No Component part Material 1 Antenna PTFE...

Page 55: ...Micropilot FMR62 Endress Hauser V 4 Rev 2 19 03 2018 55 No Component part Material 5 Process connection 316L 1 4404 PTFE cladded 6 DIN11851 slotted nut 304L 1 4307 ...

Page 56: ... cover 1 2 3 4 5 6 A0015473 No Part Material 1 Protection cover 316L 1 4404 2 Molded rubber part 4x EPDM 3 Clamping screw 316L 1 4404 carbon fibre 4 Bracket 316L 1 4404 5 Ground terminal Screw A4 Spring washer A4 Clamp 316L 1 4404 Holder 316L 1 4404 6 Washer A4 Cheese head screw A4 70 ...

Page 57: ... of the individual parameter functions Standardized operation at the device and in the operating tools Integrated data storage device HistoROM Enables transfer of configuration when changing electronic modules Records up to 100 event messages in the device Records up to 1000 measured values in the device Saves the signal curve on commissioning which can later be used as a reference Efficient diagn...

Page 58: ...lements local operation with 3 push buttons external operation via touch control 3 optical keys Operating elements also accessible in various hazardous areas Additional functionality Data backup function The device configuration can be saved in the display module Data comparison function The device configuration saved in the display module can be compared to the current device configuration Data t...

Page 59: ...ration Via HART protocol 1 4 5 7 9 6 8 2 3 A0036169 25 Options for remote operation via HART protocol 1 PLC programmable logic controller 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubox FXA191 FXA195 and Field Communicator 375 475 4 Field Communicator 475 5 Computer with operating tool e g DeviceCare FieldCare AMS Device Manager SIMATIC PDM 6 Commub...

Page 60: ...eviceCare FieldCare via service interface CDI 1 2 3 A0032466 26 DeviceCare FieldCare via service interface CDI 1 Service interface CDI of the instrument Endress Hauser Common Data Interface 2 Commubox FXA291 3 Computer with DeviceCare FieldCare operating tool ...

Page 61: ...as when levels drop below the safety stock level or plan points are indicated by the software In addition SupplyCare can also notify pre defined users by e mail Alarms If technical problems occur e g connection issues alarms are triggered and alarm e mails are sent to the System Administrator and the Local System Administrator Delivery planning The integrated delivery planning function automatical...

Page 62: ...rise SCE30B 1 SupplyCare Enterprise via Web browser 2 SupplyCare Enterprise installation 3 SupplyCare Enterprise on mobile devices via Web browser 4 Ethernet WLAN UMTS 5 Fieldgate FXA42 6 Power supply 24 V DC 7 Modbus TCP via Ethernet as server client 8 Converter from Modbus to HART Multidrop 9 HART Multidrop 10 4 x 4 to 20 mA analog input 2 wire 4 wire ...

Page 63: ...ldgate FXA42 6 Power supply 24 V DC 7 Modbus TCP via Ethernet as server client 8 Converter from Modbus to HART Multidrop 9 HART Multidrop 10 4 x 4 to 20 mA analog input 2 wire 4 wire With SupplyCare Hosting users do not need to make the initial software purchase or install and run the IT infrastructure needed Endress Hauser constantly update SupplyCare Hosting and enhance the capability of the sof...

Page 64: ...Additional safety instructions must be followed for use in hazardous areas Please refer to the separate Safety Instructions XA document included in the delivery Reference to the applicable XA can be found on the nameplate For details on the available certificates and associated XAs 83 Dual seal according to ANSI ISA 12 27 01 The devices have been designed according to ANSI ISA 12 27 01 as dual sea...

Page 65: ...hin the scope of the Pressure Equipment Directive irrespective of the maximum allowable pressure Reasons According to Article 2 point 5 of EU Directive 2014 68 EU pressure accessories are defined as devices with an operational function and having pressure bearing housings If a pressure instrument does not have a pressure bearing housing no identifiable pressure chamber of its own there is no press...

Page 66: ...5 North 11 55 35 East Switzerland Bleien 47 20 26 North 08 06 44 East Spain Yebes 40 31 27 North 03 05 22 West Robledo 40 25 38 North 04 14 57 West Hungary Penc 47 47 22 North 19 16 53 East As a general rule the requirements outlined in EN 302729 1 2 must be observed Radio standard EN 302372 1 2 The devices comply with the Tanks Level Probing Radar TLPR radio standard EN 302372 1 2 and are approve...

Page 67: ...prevent RF emissions which can otherwise interfere with aeronautical navigation The installer user of this device shall ensure that it is at least 10 km from the Dominion Astrophysical Radio Observatory DRAO near Penticton British Columbia The coordinates of the DRAO are latitude 49 19ʹ15ʹʹ N and longitude 119 37ʹ12ʹʹ W For devices not meeting this 10 km separation e g those in the Okanagan Valley...

Page 68: ... that do not have CRN approval are not listed in this table Please refer to the product structure to discover which process connections are available for a specific device type CRN approved devices bear the registration number tbd on the nameplate An additional pressure restriction applies for the device versions listed in the table below if they are CRN approved The pressure range indicated in th...

Page 69: ... JL Declaration of Conformity EC1935 2004 KE Pressure test internal procedure inspection certificate KI PMI test XRF internal procedure inspection certificate KV Declaration of Conformity ASME B31 3 The construction the material used the pressure and temperature ranges and the labeling of the devices meet the requirements of ASME B31 3 Test reports declarations and inspection certificates are avai...

Page 70: ...quirements NAMUR NE 21 Electromagnetic compatibility EMC of industrial process and laboratory control equipment NAMUR NE 43 Standardization of the signal level for the breakdown information of digital transmitters with analog output signal NAMUR NE 53 Software of field devices and signal processing devices with digital electronics NAMUR NE 107 Status classification as per NE107 NAMUR NE 131 Requir...

Page 71: ... The Configure button to the right of the product image opens the Product Configurator From your Endress Hauser Sales Center www addresses endress com Product Configurator the tool for individual product configuration Up to the minute configuration data Depending on the device Direct input of measuring point specific information such as measuring range or operating language Automatic verification ...

Page 72: ... first measuring point R Reference point of measurement 1 First measuring point 2 Second measuring point in the middle between the first and third measuring point 3 Third measuring point Measuring point Position 1st measuring point At a distance A from the reference point A length of antenna 500 mm 19 7 in Minimum distance Amin 1000 mm 39 4 in 2nd measuring point In the middle between the 1st and ...

Page 73: ... F R A F E A0032286 Minimum distance between reference point R and 100 mark Minimum span Maximum value for Empty calibration A length of antenna 200 mm 8 in Minimum value 400 mm 16 in F 400 mm 16 in E 24 m 79 ft The linearity check is performed under reference operating conditions The selected values for Empty calibration and Full calibration are only used to create the linearity protocol Afterwar...

Page 74: ...on of the measuring point marking To be selected in the additional specifications Tag plate Stainless Steel Self adhesive paper label Supplied label plate RFID TAG RFID TAG Tag plate Stainless Steel RFID TAG Self adhesive paper label RFID TAG Supplied label plate Definition of the measuring point designation To be defined in the additional specifications 3 lines containing up to 18 characters each...

Page 75: ...an asset management system e g FieldCare DeviceCare an automation system e g PLC Advantages Device condition information is available immediately and processed in time The status signals are classified in accordance with VDI VDE 2650 and NAMUR recommendation NE 107 and contain information about the cause of the error and remedial action Detailed description See Operating Instructions of the device...

Page 76: ...n be used to prove quality measures to a third party Heartbeat Verification can replace other maintenance tasks e g periodic check or extend the test intervals SIL WHG locked devices 10 The Heartbeat Verification module contains a wizard for the proof test which must be performed at appropriate intervals for the following applications SIL IEC61508 IEC61511 WHG German Water Resources Act To perform...

Page 77: ...dissolves the foam This wizard can be used via FieldCare DeviceCare or a DTM based process control system Build up detection wizard The Heartbeat Monitoring module contains the Build up detection wizard The wizard is used to configure automatic buildup detection which detects the buildup of deposits on the antenna on the basis of the increased area of the coupling signal Buildup detection can be l...

Page 78: ... 74 A0015466 300 11 8 275 10 8 255 1 10 1 5 6 6 1 4 3 8 1 5 164 6 46 54 2 13 3 5 6 5 5 5 A0015472 30 Weather protection cover Dimensions mm in The weather protection cover can be ordered together with the device product structure feature 620 Accessory Enclosed option PB Weather Protection Cover Alternatively it can be separately ordered as an accessory order code 71162242 ...

Page 79: ...HX50 display is to be retrofitted you must select version B Not prepared for display FHX50 under feature 050 Measuring device version when ordering the FHX50 In this case a retrofit kit for the device is supplied with the FHX50 The kit can be used to prepare the device so that the FHX50 can be used Use of the FHX50 may be restricted for transmitters with an approval A device can only be retrofitte...

Page 80: ...28617 For 2 channel devices feature 020 option B C E or G OVP20 71128619 Hosuing lid for retrofitting In order to keep the necessary safety distances the housing lid needs to be replaced if the device is retrofitted with the overvoltage protection Depending on the housing type the order code of the suitable lid is as follows GT18 housing Lid 71185516 GT19 housing Lid 71185518 GT20 housing Lid 7118...

Page 81: ...e applications with SupplyCare Hosting Up to 4 field devices with 4 to 20 mA communication FXA30 FXA30B serial Modbus FXA30B or HART FXA30B can be connected With its robust design and ability to run for years on the battery it is ideal for remote monitoring in isolated locations Version with LTE USA Canada and Mexico only or 3G mobile transmission for worldwide communication For details see Techni...

Page 82: ...ION Fieldbus devices Technical Information TI01134S DeviceCare is available for download at www software products endress com The download requires a registration in the Endress Hauser software portal Alternatively a DeviceCare DVD can be ordered with the device Product structure Feature 570 Service Option IV Tooling DVD DeviceCare Setup FieldCare SFE500 FDT based Plant Asset Management tool Helps...

Page 83: ...a Gb XA01549F XA01549F XA01549F BC ATEX II 1 2G Ex ia db ia Ga IIC T6 Ga Gb XA01552F XA01552F XA01552F BD ATEX II 1 2 3G Ex ia ic ia Ga IIC T6 Ga Gb Gc XA01550F XA01550F XA01550F BG ATEX II 3G Ex ec IIC T6 Gc XA01551F XA01551F XA01551F BH ATEX II 3G Ex ic IIC T6 Gc XA01551F XA01551F XA01551F BL ATEX II 1 2 3G Ex ia ec ia Ga IIC T6 Ga Gb Gc XA01550F XA01550F XA01550F B2 ATEX II 1 2G Ex ia IIC T6 Ga...

Page 84: ...JPN Ex ia IIIC T85 C Da Db XA01728F 4 XA01728F 4 XA01728F 4 J3 JPN Ex d ia IIC T6 Ga Gb JPN Ex ta tb IIIC T85 C Da Db XA01729F 4 XA01729F 4 XA01729F 4 J4 JPN Ex ia IIC T6 Ga Gb JPN Ex d ia IIC T6 Ga Gb XA01726F 4 XA01726F 4 XA01726F 4 KA KC Ex ia IIC T6 Ga XA01623F 4 XA01623F 4 XA01623F 4 KB KC Ex ia IIC T6 Ga Gb XA01623F 4 XA01623F 4 XA01623F 4 KC KC Ex ia db ia Ga IIC T6 Ga Gb XA01624F 4 XA01624...

Page 85: ......

Page 86: ......

Page 87: ......

Page 88: ...www addresses endress com 71394729 71394729 ...