Operation
Memosens Wave CAS80E
28
Hauser
1. Select data record.
2. Set the calibration point in the medium and enter the target sample value (laboratory
value).
2-point calibration
Measured value deviations are to be compensated for at 2 different points in an
application (e.g. the maximum and minimum value of the application). This aims to ensure
a maximum level of accuracy between these two extreme values.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
x
y
a
b
A0039325
25
Principle of a 2-point calibration
x
Measured value
y
Target sample value
a
Factory calibration
b
Application calibration
1. Select a data record.
2. Set 2 different calibration points in the medium and enter the corresponding set
points.
A linear extrapolation is performed outside the calibrated operational range (gray
line).
The calibration curve must be monotonically increasing.