Mounting
Liquicap M FMI52 HART
14
Hauser
4.2.3
Wall mounting
1 →
2 →
3
A0042318
A0042319
A0042320
‣
Screw together the wall bracket
on the tube.
‣
Mark the distance between the
holes on the wall before drilling.
‣
Screw the separate housing on
the wall.
4.2.4
Pipe mounting
The maximum pipe diameter is 50.8 mm (2 in).
1 →
2
A0042318
A0042321
‣
Screw together the wall bracket
on the tube.
‣
Screw the separate housing on a
pipe.
4.2.5
Shortening the connecting cable
NOTICE
Risk of damage to connections and cable.
‣
Make sure that neither the connecting cable nor the probe is turning with the pressing
screw!
Recalibration must be performed before commissioning.
The maximum connection length between the probe and the separate housing is
6 m (20 ft).
When ordering a device with separate housing, the desired length must be specified.