Levelflex M FMP43 with PROFIBUS PA
Wiring
Hauser
27
Wiring in T12 housing
L00-FMP41Cxx-04-00-00-en-004
-
-
1 2
3 4
T-Box
5
3 4
1 2
- +
1
"
Ser.-No.:
Order Code:
D00886-A
t >85°C
x =
if modification
see sep. label
Dat./Insp.:
PN max.
TAntenne max.
°C
79689
Maulburg
Made
in
German
y
T >70°C :
A
LEVELFLEX M
HAUSER
Profibus PA
PTB 00 ATEX
II 1/2 G EEx ia IIC T6
IP65
x x x x x x x x
x x x x x x x x
1
3
4
11
8
7
2
4
3
7
8
When you use the measuring system in hazardous areas, make sure you comply with
national standards and the specifications in the safety instructions (XA’s).
Make sure you use the specific cable gland.
Connect up the Levelflex M as follows:
Before unscrew housing cover (2) at seperate connection room
turn off the power supply!
Insert cable (3) through gland (5).
Use screened, twisted wire pair.
Only ground screening of the line (5) on sensor side.
Make connection (see pin assignment).
Tighten cable gland (4).
Screw on housing cover (2).
Switch on power supply.
l
l
l
l
l
plant
ground
Before connection please note the following:
PROFIBUS devices are marked on the nameplate (1). The
voltage is determined by the PROFIBUS standard and the
desired safety concept. (see chapter 4.3).
Connect potential matching line to transmitter earth terminal
before connecting up the device.
Tighten the locking screw:
It forms the connection between the probe and the housing
earth potential.
l
l
l
Caution!
Summary of Contents for Levelflex M FMP43
Page 4: ...Brief overview Levelflex M FMP43 with PROFIBUS PA 4 Endress Hauser...
Page 104: ...Appendix Levelflex M FMP43 with PROFIBUS PA 104 Endress Hauser...
Page 106: ...Levelflex M FMP43 with PROFIBUS PA Index 106 Endress Hauser...
Page 108: ...www endress com worldwide BA358F 00 en 07 07 No 71041163 CCS FM SGML 6 0 ProMoDo 71041163...