Levelflex FMP55 HART
Mounting
Hauser
29
Vessels with heat insulation
If process temperatures are high, the device must be included in normal tank
insulation to prevent the electronics heating up as a result of heat radiation or
convection. The insulation may not exceed beyond the points labeled "MAX" in the
drawings.
40(1.57)
40(1.57)
MAX
mm (in)
MAX
1
3
2
A0014654
5
Process connection with flange - FMP55
1
Tank insulation
2
Compact device
3
Sensor remote (feature 600)