Proline t-mass I 300 HART
Diagnostics and troubleshooting
Hauser
143
11.5 Diagnostic information in FieldCare or DeviceCare
11.5.1 Diagnostic options
Any faults detected by the measuring device are displayed on the home page of the
operating tool once the connection has been established.
Xxxxxx/…/…/
Xxxxxx
P
P
P
+
+
+
+
Diagnostics 1:
Remedy information:
Failure (F)
Function check (C)
Out of spezification (S)
Maintenance required (M)
C485 Simulation measured vari...
Deactivate Simulation (Service...
Diagnostics 1:
C485 Simu...
Remedy information:
Deactivate...
Access status tooling:
Mainenance
Operation
Setup
Diagnostics
Expert
Function check (C)
Status signal:
Device tag:
Device name:
Mass flow:
Volume flow:
Xxxxxxx
12.34
kg/h
12.34
m /h
³
Xxxxxxx
1
2
3
A0021799-EN
1
Status area with status signal→ 139
2
3
Remedy information with Service ID
In addition, diagnostic events which have occurred can be shown in the
Diagnostics
menu:
• Via parameter → 148
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault. In addition, the corresponding symbol for the
diagnostic behavior is displayed in front of the diagnostic information on the local display.
Diagnostic information
Diagnostic code
Diagnostic
behavior
Status signal
Diagnostic
number
Short text
↓
↓
↓
Example
A0013962
A0013958
441
Curr.output 1
NAMUR
NE 107
3-digit number