Commissioning
Proline t-mass I 300 HART
122
Hauser
Parameter
Prerequisite
Description
Selection / User entry
Pulse output simulation 1 to n
In the
Operating mode
parameter, the
Pulse
option is selected.
Set and switch off the pulse output
simulation.
For
Fixed value
option:
Pulse
width
parameter (→ 93)
defines the pulse width of the
pulses output.
• Off
• Fixed value
• Down-counting value
Pulse value 1 to n
In the
Pulse output simulation 1 to n
parameter, the
Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65
535
Switch output simulation 1 to n
In the
Operating mode
parameter, the
Switch
option is selected.
Switch the simulation of the switch
output on and off.
• Off
• On
Switch status 1 to n
–
Select the status of the status output for
the simulation.
• Open
• Closed
Relay output 1 to n simulation
–
Switch simulation of the relay output
on and off.
• Off
• On
Switch status 1 to n
The
On
option is selected in the
Switch
output simulation 1 to n
parameter
parameter.
Select status of the relay output for the
simulation.
• Open
• Closed
Device alarm simulation
–
Switch the device alarm on and off.
• Off
• On
Diagnostic event category
–
Select a diagnostic event category.
• Sensor
• Electronics
• Configuration
• Process
Diagnostic event simulation
–
Select a diagnostic event to simulate
this event.
• Off
• Diagnostic event picklist
(depends on the category
selected)
*
Visibility depends on order options or device settings
9.8
Protecting settings from unauthorized access
The following write protection options exist in order to protect the configuration of the
measuring device from unintentional modification:
• Protect access to parameters via access code → 122
• Protect access to local operation via key locking → 58
• Protect access to measuring device via write protection switch → 124
9.8.1
Write protection via access code
The effects of the user-specific access code are as follows:
• Via local operation, the parameters for the measuring device configuration are write-
protected and their values can no longer be changed.
• Device access is protected via the Web browser, as are the parameters for the measuring
device configuration.
• Device access is protected via FieldCare or DeviceCare (via CDI-RJ45 service interface), as
are the parameters for the measuring device configuration.
Defining the access code via local display
1. Navigate to the
Define access code
parameter (→ 112).
2. Define a max. 16-digit character string comprising numbers, letters and special
characters as the access code.