Source container with source insert for manual or pneumatic switch-ON/switch-

OFF

Application

The FQG61 and FQG62 source containers are designed to hold the radioactive source

during radiometric point level measurement, continuous level measurement and

density measurement. The radiation is emitted almost unattenuated in one direction

only, and is damped in all other directions. FQG61 and FQG62 differ from each other

in terms of size and screening effect.

Your benefits

• Lightweight device provides best possible screening thanks to almost spherical

design

• Safe and easy source replacement

• Highest safety classification for the source supplied (DIN 25426/ISO 2919, typical

classification C66646)

• Compact device that is easy to mount

• Various angles of emission for optimum adaptation to the application

• Manual or pneumatic switch-ON/switch-OFF

• Padlock, cylinder lock or locking bolt for fixing the switch position

• Switch state easily identified

• Fire-resistant v821 °C (+1510 °F) / 30 minutes

Products

Solutions

Services

Technical Information

Source container FQG61,

FQG62

Radiometric level measurement

Technical Information and Operating Instructions

TI00435F/00/EN/19.19

71456915

2019-10-22

Summary of Contents for FQG61 Series

Page 1: ...des best possible screening thanks to almost spherical design Safe and easy source replacement Highest safety classification for the source supplied DIN 25426 ISO 2919 typical classification C66646 Co...

Page 2: ...ns 33 Orientation for level measurement 34 Orientation for point level detection 35 Orientation for density measurement 36 Orientation of the fireproof version 36 Mounting device supplied by customer...

Page 3: ...iner FQG61 FQG62 Endress Hauser 3 Product identification Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3...

Page 4: ...res and other facts which do not result in personal injury Symbols for certain types of information Symbol Meaning Permitted Procedures processes or actions that are permitted Preferred Procedures pro...

Page 5: ...uctions for loading and changing the source Label set Clamping device FHG61 Documentation Comment SD01221F 00 Clamping device FHG61 For rectangular and diagonally irradiated pipes with a diameter 50 t...

Page 6: ...FTG20 Documentation Comment TI01023F 00 Technical Information for Gammapilot FTG20 BA01035F 00 Operating Instructions for Gammapilot FTG20 RFID TAG Documentation Comment SD01502F 00 document which is...

Page 7: ...Source container FQG61 FQG62 Endress Hauser 7 Certificate of suitability A0037353...

Page 8: ...k Before switching on the radiation make sure that no one is in the radiation zone or inside the product vessel The radiation may only be switched ON by properly instructed staff Do not operate damage...

Page 9: ...ssible The local radiation dose rate decreases in proportion to the square of the distance from the radiation source Legal regulations for radiation protection The handling of radioactive emitters is...

Page 10: ...urther information on radiation protection and regulations in other countries Supplementary safety instructions Observe the associated safety instructions contained in the following documents SD00292F...

Page 11: ...of the radiation source containers is available product structure order code 670 Additional function This version has a compensation compartment which is welded onto the housing In the event of a fir...

Page 12: ...and column Configure the selected product The Product Con gurator for the selected product opens Dose rate diagrams for 60Co 90 90 30 30 60 60 120 120 300 300 330 330 0 0 270 270 210 210 240 240 180 1...

Page 13: ...FQG62 Activity in MBq FQG61 Max value 100 in Sv h FQG62 Max value 100 in Sv h AA 3 7 3 7 0 01 0 01 AB 7 4 7 4 0 01 0 01 AC 18 5 18 5 0 01 0 01 AD 37 37 0 01 0 01 AE 74 74 0 02 0 01 AF 111 111 0 04 0...

Page 14: ...st and humidity Swivel bracket for manual switch ON switch OFF Padlock to secure the ON OFF switch position K L Pneumatic switch ON switch OFF Padlock to secure the OFF switch position M N Higher prot...

Page 15: ...FQG61 FQG62 order code 020 option B 54 B C 20 0 79 220 8 66 A A0018386 2 Dimensions mm in Dimensions Version mm in Comment A FQG61 251 9 88 FQG62 272 10 7 B FQG61 287 11 3 FQG62 368 14 5 C FQG61 450...

Page 16: ...7 Clearance required to replace the radiation source FQG62 570 22 4 FQG61 FQG62 order code 020 option D 54 B C 20 0 79 220 8 66 A 1 A0018388 4 Dimensions mm in 1 Reference O ring Dimensions Version mm...

Page 17: ...rence O ring Dimensions Version mm in Comment A FQG61 251 9 88 FQG62 272 10 7 B FQG61 427 16 8 FQG62 508 20 0 C FQG61 483 19 0 Clearance required to replace the radiation source FQG62 602 23 7 Fire re...

Page 18: ...e of the radiation source container The radiation emission channel is indicated by a mark on the covering plate of the mounting flange Angle of emission According to feature 240 of the product structu...

Page 19: ...71 S235JR 1 0038 Flange 316L 1 4404 2 Housing ring 316L 1 4404 304 1 4301 3 Nameplate 316L 1 4404 4 Cover 304 1 4301 O ring FKM 5 Screw groove pin A2 6 Warning sign Acrylate foil 7 Nameplate for radia...

Page 20: ...AUS OFF sign 304 1 4301 5 Source nameplate 304 1 4301 6 Rotary element 316L 1 4404 7 Screw A4 8 Screw A4 Nut A4 9 Bracket A4 10 Padlock body Brass Padlock shackle Hardened steel 11 EIN ON sign 304 1 4...

Page 21: ...ackle Hardened steel 3 Indication plate 316L 1 4404 4 AUS OFF sign 304 1 4301 5 Rotary element 316L 1 4404 6 Source nameplate 304 1 4301 7 Screw A4 8 Swivel bracket 316L 1 4404 9 EIN ON sign 304 1 430...

Page 22: ...AUS OFF sign 304 1 4301 5 Swivel bracket 316L 1 4404 6 Source nameplate 304 1 4301 7 Additional national sign 304 1 4301 8 Mounting A2 9 Padlock body Brass Padlock shackle Hardened steel 10 Screw A4 S...

Page 23: ...il 4 Caution radiation sign 304 1 4301 5 6 Tag 304 1 4301 7 Radioactive material sign 304 1 4301 8 Padlock body Brass Padlock shackle Hardened steel 9 Securing plate 316L 1 4404 10 Terminal housing PC...

Page 24: ...ion ensure The ON or OFF switch position is secured Protection against theft Pneumatic drive The following applies for the version with pneumatic switch ON switch OFF Swivel range 180 Compressed air c...

Page 25: ...tag is used temperature range restrictions must be taken into consideration See SD01502F 00 Ambient pressure Atmospheric pressure Vibration resistance DIN EN 60068 2 64 test Fh 10 to 2000 Hz 1 g2 Hz F...

Page 26: ...abeling of OFF switch position plus additional language German French Swedish Norwegian Russian 7 Labeling of ON switch position plus additional language German French Swedish Norwegian Russian 8 Inte...

Page 27: ...Order code of radiation source container according to product structure 54 4 Order code of radiation source container according to product structure 54 5 Radiation angle of emission 6 Local dose rate...

Page 28: ...5 4 3 2 1 6 9 8 Endress Hauser Australia Pty Ltd Level 1 16 Giffnock Avenue NORTH RYDE NSW 2113 Telephone 1800 363 7377 ENDRESS HAUSER CAUTION RADIATION SOURCE MANUFACTURED BY DATE OF MANUFACTURE SOUR...

Page 29: ...Serial number of the source capsule according to certificate 6 Activity including unit MBq or GBq 7 Date of loading month year Additional nameplate of radiation source A A Made in Germany D 79689 Mau...

Page 30: ...fferent from the local dose rate which has been calculated from the specified attenuation factors 11 RFID TAG Operating principle of RFID and NFC Radio frequency identification RFID makes it possible...

Page 31: ...Mobile device that supports NFC 2 RFID TAG The RFID TAGs of the radiation source FSG60 FSG61 and the radiation source container FQG61 FQG62 are identical in appearance The only difference is the data...

Page 32: ...or 380 x 380 x 600 mm 15 x 15 x 23 6 in NOTICE The foam packaging can be disposed of as normal household waste NOTICE The radioactive warning labels triangular must not be removed All other labels can...

Page 33: ...fied specially trained personnel whose radiation exposure is monitored Ensure that this is allowed by the handling permit All local conditions must be taken into consideration All work should be carri...

Page 34: ...nted on the opposite side The source container and detector should be mounted as close as possible to the product vessel to avoid control zones a 2 a 2 1 A0018401 1 FQG61 FQG62 2 FMG60 Angle of emissi...

Page 35: ...r accuracy reasons 2 1 A0018403 1 FQG61 FQG62 2 FMG60 Orientation for point level detection The version of the source container with an angle of emission of 5 is recommended for point level detection...

Page 36: ...nting the radiation source container and the FMG60 compact transmitter on pipes Clamping device FHG61 55 Measurement section FHG62 56 Orientation of the fireproof version Orientation I recommended The...

Page 37: ...A Example of a mounting plate B Example for L profiles Dimensions EN ANSI A 68 9 mm 2 71 in 72 9 mm 2 87 in B 166 3 mm 6 55 in 176 0 mm 6 93 in C 180 0 mm 7 09 in 190 5 mm 7 5 in D 18 0 mm 0 71 in 19...

Page 38: ...d once the radiation source has been installed LCAUTION Depending on the installation radiation can also occur outside the actual radiation emission channel through scattering In such cases it must be...

Page 39: ...ion 5 Vent filter 6 Connecting cable of proximity switch for AUS OFF switch position 7 Connecting cable of proximity switch for EIN ON switch position 8 Proximity switch for EIN ON switch position 9 P...

Page 40: ...commissioning the compressed air supply must be connected for the pneumatic drive and the padlock 1 on the top must be removed This padlock only needs to be affixed again for revision purposes OFF po...

Page 41: ...surized Technical data of the pneumatic actuator Swivel range 180 Compressed air connection G1 8 Actuating pressure 3 5 to 6 bar 51 to 87 psi Reset by spring force Required compressed air quality ISO...

Page 42: ...rn the cover approx 45 counterclockwise as far as it will go 2 Remove the cover 1 2 A0018415 3 Using the key turn the closing cylinder approx 45 in the counterclockwise direction 3 A0033938 4 Pull out...

Page 43: ...ing the switch state Radiation ON the EIN ON sign is visible The arrow points to EIN ON Radiation OFF the AUS OFF sign is visible The arrow points to AUS OFF FQG61 FQG62 order code 020 option B Switch...

Page 44: ...n that is visible EIN ON or AUS OFF The other sign is covered by the swivel bracket EIN ON AUS OFF 4 A0018417 5 Fix the padlock in the position provided 6 Let the locking bolt snap into place in the E...

Page 45: ...itching ON the radiation LWARNING Rotating the insert over the locking pin leads to the removal position of the source insert Do not press the sealed locking pin A 1 Remove the padlock 2 Release the l...

Page 46: ...ructions for switching ON the radiation Before switching on the radiation make sure that no one is in the radiation zone or inside the product vessel The radiation may only be switched ON by properly...

Page 47: ...dicated by the sign that is visible EIN ON or AUS OFF The other sign is covered by the swivel bracket 3 A0018423 4 Secure the ON switch position by hooking the padlock into the position provided 5 Tig...

Page 48: ...s of corrosion at the source container the local dose rate around the device must be measured If the value is significantly above the normal operation levels cordon off the area and notify the radiati...

Page 49: ...t only required as part of routine checks but must also be performed whenever an incident occurs that may impair the casing around the radiation source In such instances the leak test procedure must b...

Page 50: ...020 option K L M N along the thread of the proximity switches and the three annular grooves on the cylinder housing 2 Have the samples analyzed by an authorized organization A source is considered to...

Page 51: ...nbolted from its mounting position Point the radiation emission channel towards a thick wall e g steel or lead or mount a blind flange in front of the emission channel Individuals should at all times...

Page 52: ...many FRA Conditions The following conditions must be met before returning the device An inspection certificate no more than three months old and confirming the leak tightness of the radiation source m...

Page 53: ...rtificate of the source is enclosed with the return shipment documents No significant signs of corrosion on the container which could put the safe storage of the source at risk No signs of serious dam...

Page 54: ...ting Instructions SD00293F 00 for delivery to the United States Safety Instructions ATEX II 2 G XA01633F 00 Delivery Germany We can only ship radioactive sources once we have received a copy of the ha...

Page 55: ...he product picklists measurement method product family etc Device support right hand column Configure the selected product The Product Con gurator for the selected product opens at your Endress Hauser...

Page 56: ...product family etc Device support right hand column Configure the selected product The Product Con gurator for the selected product opens at your Endress Hauser Sales Center www addresses endress com...

Reviews: