Flowfit CUA252
Maintenance
Hauser
15
5
Maintenance
You must perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or
calendar.
The maintenance cycle primarily depends on the following:
• The facility
• The installation conditions
• The medium being measured
CAUTION
!
Risk of injury if medium escapes
‣
Prior to performing any maintenance task, make sure that the process pipe is
unpressurized, empty and rinsed.
5.1
Cleaning the assembly
• Use suitable cleaning solutions (see "Cleaning agents" section) to remove light dirt and
fouling.
• Heavy dirt can be removed with a soft brush and a suitable cleaning agent.
• To remove stubborn dirt, soak the parts in a cleaning solution. Then clean the parts with a
brush.
Six months is a typical cleaning interval for drinking water, for example.
5.2
Cleaning the sensor
Cleaning with the ultrasonic cleaning system
If the (optional) CYR52 ultrasonic cleaning system is installed, you can clean the sensor
surfaces during operation.
Enter the following parameters to prevent the ultrasonic transducer from overheating:
Cleaning without the ultrasonic cleaning system
If the assembly is not fitted with an ultrasonic cleaning system, you must remove the sensor
in order to clean it.
You must rinse the sensor thoroughly with water after cleaning it.
Cleaning time:
max. 5 seconds
Cleaning interval:
min. 5 minutes
Type of fouling
Cleaning measure
Lime deposits
Immerse the sensor in 1%-5 % hydrochloric acid (for a few minutes).
Dirt particles on the optical windows
Use a cleaning cloth to clean the optical windows.