Installation
Dosimass
8
Hauser
3.2.2
Mounting location
Correct measurement is only possible if the pipe is filled.
For this reason, avoid
the following
mounting locations in the pipe:
• At the highest point of the pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a down pipe.
A0008566
Fig. 2:
Mounting location
The following proposed installation, however, permits installation in an open down pipe. Pipe
restrictors or the use of an orifice with a cross-section smaller than the nominal diameter prevent
the pipe from running empty during measurement.
A0008565
Fig. 3:
Installation in a down pipe (e.g. for batching applications)
1
Supply tank
2
Sensor
3
Orifice plate, pipe restriction
4
Valve
5
Batching tank
1
2
3
4
5
Dosimass / DN
8 (3/8")
15 (½")
25 (1")
Ø Orifice plate, pipe restriction
6 mm (0.24 in)
10 mm (0.39 in)
14 mm (0.55 in)