Dosimass
Appendix - Function description
Hauser
53
11.10
Function group SIMULATION
11.11
Function group SENSOR VERSION
11.12
Function group AMPLIFIER VERSION
Function description
Function group SIMULATION
SIMULATION
MEASURAND
Use this function to simulate a measured variable to check that the pulse output behaves
correctly, for example. During this time, the words "SIMULATION MEASURAND"
appear in the operating program.
Options:
OFF
MASS FLOW
VOLUME FLOW
DENSITY
TEMPERATURE
Factory setting:
OFF
"
Caution!
• The measuring device cannot be used for measuring while this simulation is in
progress.
• The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
!
Note!
This function is not available unless the SIMULATION MEASURAND function was
activated.
Use this function to define a selectable value (e.g. 12 kg/s). This value is used to test
downstream devices and the measuring device itself.
Factory setting:
0 kg/h or 0 m
3
/h or 0 kg/l or 0°C
"
Caution!
The setting is not saved if the power supply fails.
Function description
SENSOR VERSION function group
SERIAL NUMBER
Use this function to display the serial number of the device.
SENSOR TYPE
Use this function to display the sensor type.
SOFTWARE REV. DAT
Use this function to display the revision number of the software used to program the
DAT.
Function description
AMPLIFIER VERSION function group
SOFTWARE REV.
AMPLIFIER
Use this function to display the software revision number of the amplifier.