Dosimag
Diagnostics and troubleshooting
Hauser
49
Diagnostic
number
Short text
Remedy instructions
Status signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
834
Process temperature
too high
Reduce process temperature
S
Warning
835
Process temperature
too low
Increase process temperature
S
Warning
937
EMC interference
Change main electronic module
S
Warning
938
EMC interference
1. Check ambient conditions
regarding EMC influence
2. Change main electronic
module
F
Alarm
991
Batch time exceeded
Check process conditions
F
Warning
991
Maximum flow rate
exceeded
F
Warning
1)
Diagnostic behavior can be changed.
12.5 Pending diagnostic events
The
Diagnostics
menu provides the option of displaying the current and previous
diagnostic event separately.
To call up the measures to rectify a diagnostic event:
Via the DeviceCare and FieldCare operating tool→ 47
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Structure of the submenu
Diagnostics
→
Actual diagnostics
Previous diagnostics
Parameter overview with brief description
Parameter
Prerequsite
Description
User interface
Actual diagnostics
A diagnostic event has occurred.
Shows the current occured diagnostic
event along with its diagnostic
information.
If two or more messages occur
simultaneously, the message with
the highest priority is shown on
the display.
Symbol for diagnostic
behavior, diagnostic code
and short message.
Previous diagnostics
Two diagnostic events have already
occurred.
Shows the diagnostic event that
occurred prior to the current diagnostic
event along with its diagnostic
information.
Symbol for diagnostic
behavior, diagnostic code
and short message.