Dosimag
Commissioning
Hauser
35
Parameter
Prerequsite
Description
Selection / User
entry
Factory setting
Value per pulse
One of the following options is
selected in the
Operating
mode
parameter:
• Pulse
• Automatic pulse
In the
Assign pulse output
Volume flow
option is
selected.
Enter measured value at which
a pulse is output.
Signed floating-point
number
Depends on the
nominal diameter:
• DN 4 (¹/₈"):
0.005 ml
(0.0002 fl oz)
• DN 8 (³/₈"):
0.02 ml
(0.001 fl oz)
• DN 15 (½"):
0.1 ml
(0.004 fl oz)
• DN 15K (½K"):
0.1 ml
(0.004 fl oz)
• DN 25 (1"):
0.2 ml
(0.007 fl oz)
Pulse width
In the
Operating mode
parameter, the
Pulse
option is
selected and in the
Assign
pulse output
parameter
(→ 34), the
Volume flow
option is selected.
Define time width of the
output pulse.
0.05 to 3.75 ms
0.05 ms
Failure mode
One of the following options is
selected in the
Operating
mode
parameter:
• Pulse
• Automatic pulse
In the
Assign pulse output
Volume flow
option is
selected.
Define output behavior in
alarm condition.
• Actual value
• No pulses
Actual value
Invert output signal
–
Invert the output signal.
• No
• Yes
• Pulse/freq./switch
output 1: yes
• Pulse/freq./switch
output 2: no
Configuring the frequency output
Flow-proportional frequency output with on/off ratio of 1:1
Navigation
"Setup" menu → Pulse/frequency/switch output 1 to 2
Structure of the submenu for the frequency output
‣
Pulse/frequency/switch output 1
to 2
Operating mode
Assign frequency output
Minimum frequency value
Maximum frequency value