Basic safety instructions
Condumax CLS15/16/21
6
Hauser
Temperature classes
Name
Type
Medium temp. T
a
for temperature class (Tn)
Cat.
x1 x2 x3 x4
Condumax
CLS15
-
*
** *
A
-20 °C ≤ Ta ≤ +140 °C (T3)
-20 °C ≤ Ta ≤ +115 °C (T4)
-20 °C ≤ Ta ≤ +65 °C (T6)
II 1G
Condumax
CLS16
-
X
** *
A/B -5 °C ≤ Ta ≤ +150 °C (T3)
-5 °C ≤ Ta ≤ +115 °C (T4)
-5 °C ≤ Ta ≤ +65 °C (T6)
II 1G
Condumax
CLS21
-
*
** *
D
-20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +130 °C (T4)
-20 °C ≤ Ta ≤ +80 °C (T6)
II 1G
CLS21
-
*
** *
A
-20 °C ≤ Ta ≤ +135 °C (T3)
-20 °C ≤ Ta ≤ +115 °C (T4)
-20 °C ≤ Ta ≤ +65 °C (T6)
II 1G
X ... Variant not applicable
x1 ... Measuring range and cell constant (no Ex relevance)
x2 ... Process connection / material (no Ex relevance)
x3 ... Measuring cable connection
x4 ... Temperature sensor: A = Pt 100, B = Pt 1000, D = without temperature sensor
• If the specified medium temperatures are complied with, temperatures that are not
permitted for the respective temperature class will not occur on the equipment.
• With the exception of sensor version CLS15-*1M**, for functional reasons the CLS15
sensors may only be operated up to 120 °C (248 °F) during continuous operation / and up to
140 °C (284 °F) for short periods. The CLS15-*1M** version may only be operated up to
100 °C (212 °F) during continuous operation.
• For functional reasons, the CLS16 sensors may only be operated up to 120 °C (248 °F)
during continuous operation / and up to 150 °C (302 °F) for short periods.
The following connection values are safety limits which may not be exceeded when connecting
to the transmitter:
Parameters
Connection data
Supply circuit
Intrinsically safe
Maximum input voltage U
i
15 V
Maximum input current I
i
30 mA
Maximum input power P
i
130 mW
Maximum internal capacitance C
i
Negligible
Maximum internal inductance L
i
Negligible