Condumax CLS15/16/21
Technical data
Hauser
23
9.3
Environment
9.3.1
Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
9.3.2
Storage temperature
-25 to +80 °C (-10 to +180 °F)
9.3.3
Degree of protection
CLS15
IP 67 / NEMA 6
CLS16
Fixed cable version
IP 67 / NEMA 6
TOP68 plug-in system
IP 68 / NEMA 6
CLS21
Fixed cable version
IP 67 / NEMA 6
Plug-in head version
IP 65 / NEMA 4X
9.4
Process
9.4.1
Process temperature
CLS15
Threaded version with fixed cable
-20 to 100 °C (-4 to 212 °F)
Threaded version with plug-in head, Clamp version
Normal operation
-20 to 120 °C (-4 to 248 °F)
Sterilization (max. 1 h)
1)
Max. 140 °C (284 °F)
CLS16
Normal operation
-5 to 120 °C (23 to 248 °F)
Sterilization (max. 45 min)
Max. 150 °C (302 °F) at 6 bar (87 psi) absolute
CLS21
Threaded version with fixed cable
-20 to 100 °C (-4 to 212 °F)
Version with plug-in head, Clamp version-20 to +135 °C (-4 to 275 °F) at 3.5 bar (50 psi)
absolute
1)
Threaded versions: max. 30 minutes