Ceraphant T PTC31, PTP31, PTP35
Operation
Hauser
15
OUT/OUT2
Output/output 2
Switching
characteristic
HYNO:
Hysteresis/NO contact
HYNC:
Hysteresis/NC contact
WINO:
Window/NO contact
WINC:
Window/NC contact
4...20 mA:
Analog output (only if
available)
Switch point value
Switch point 0.5...100
% URL
in increments of 0.1 %
of selected unit (min.
0.001 bar)
Switch-back point
value
Switch-back point
0...99.5 % URL
in increments of 0.1 %
of selected unit (min.
0.001 bar)
Switch point delay
Delay time
0...99 s
in increments of 0.1 s
Switch-back point
delay
Delay time
0...99 s
in increments of 0.1 s
Note
Adjustment of measuring ranges with negative gauge pressure up to 4 bar in increments of min.
0.01 mbar
Min. distance between SP and RSP: 0.5% URL